Existing paradigms can be loaded via the
Load
button. Progress can also be saved at any time by hitting
the F1 key on the keyboard for a quick
Save
.
Configuring an olfactory condition
Navigate to the
Add / edit
valve tab to add an olfactory stimulus. As defined in the prerequisites, a
stimulus in this paradigm will be six seconds long, followed by a 30 second odorless block. The base unit
of each input block is in milliseconds. Thus, in the duration of the
valve open
period is set to 6000 and
the
valve close
duration is set to 30000 (Figure 39).
Figure 39 - Configuration of olfactory stimuli via the valve tab. This can be used to add and edit valves within a group.
The paradigm already contains a predefined device with six possible stimulus channels. If the user were
to connect an ETT Olfactometer C or similar device with more possible channels this number will
automatically adjust.
Valve 1
is already preselected as the active channel. To the right of this tab, select
the
Group settings
tab. The empty paradigm already contains a pre-defined empty group called
Group 1
Figure 40 - The group settings tab of an empty pre-defined group.
By clicking on the
Add to group
button in the valve configuration tab, the olfactory stimulus will be
added to the selected group (Figure 41).
Figure 41 - Group 1 with added olfactory stimulus
Summary of Contents for Olfactometer 2
Page 1: ...ETT Olfactometer ETTDirectControl The First Paradigm Manual and Tutorial Revision 2021 1 ...
Page 2: ...2 Emerging Tech Trans LLC ...
Page 53: ...53 Emerging Tech Trans LLC Typical air flow configuations ...
Page 54: ...54 Emerging Tech Trans LLC ...
Page 55: ...55 Emerging Tech Trans LLC Configuration for multiple channels at once ...
Page 56: ...56 Emerging Tech Trans LLC Setting up multiple channels within ETTDirectControl ...