I. Important Safety Information 

 

SAFETY INSTRUCTIONS 

The equipment described in this manual is designed and manufactured to operate within defined design 
limits. Any misuse may result in electric shock or fire. To prevent the equipment from being damaged, 
the  following  rules  should  be  observed  for  installation,  use  and  maintenance.

  Read  the  following 

safety instructions before operating the instrument. 

 

 

POWER 

 

POWER CORD: 

Use only the power cord specified for this equipment and certified for the country of use. If the 

power (mains) plug is replaced, follow the wiring connections specified for the country of use. When installing or 
removing the power plug, 

hold the plug, not the cord. 

 

The power cord provided is equipped with a

 3-prong grounded plug (a plug with a third grounding pin). 

This 

is both a safety feature to avoid electrical shock and a requirement for correct equipment operation. If the outlet 
to be used does not accommodate the 3-prong plug, either change the outlet or use a grounding adapter. 

 

FUSES: 

Replace fuses only with those having the required current rating, voltage, and specified type such as 

normal blow, time delay, etc. 

DO NOT

 use makeshift fuses or short the fuse holder. This could cause a shock or 

fire hazard or severely damage the instrument. 

 

OPERATION 

CAUTION 

DO NOT OPERATE WITH COVERS OR PANELS REMOVED. 

Voltages inside the equipment consist of 

line (mains) that can be anywhere from 

100-240VAC. 

DO NOT OPERATE WITH SUSPECTED EQUIPMENT FAILURES. 

If any odor or smoke becomes apparent 

turn off the equipment and unplug it immediately. Failure to do so may result in electrical shock, fire, or 
permanent damage to the equipment. Contact the factory for further instructions.  

DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE: 

Operating the equipment in the presence of 

flammable gases or fumes 

constitutes a definite safety hazard

. For equipment designed to operate in such 

environments the proper safety devices must be used such as dry air or inert gas purge, intrinsic safe barriers 
and/or explosion-proof enclosures. 

USE DISTILLED OR DEIONIZED WATER SOURCE FOR HUMIDIFICATION.

  Build-up of contaminates on 

the transducer will cause stress to the transducer and electronics and resulting in premature failure and 
invalidate the warranty. 

 

 

Summary of Contents for 5482

Page 1: ...www electrotechsystems com 833 ENV GURU 833 368 4878 D01016 Revision D 20210107 Humidification System Model 5482 Operating Manual...

Page 2: ...II Warranty 14 Revision History Revision A Released 2019 04 01 Initial release Revision B Released 2019 07 22 QOL additions Functionality section renamed to Operation Guide Section numbers corrected R...

Page 3: ...or use a grounding adapter FUSES Replace fuses only with those having the required current rating voltage and specified type such as normal blow time delay etc DO NOT use makeshift fuses or short the...

Page 4: ...ndard Optionally the humidification system can be connected directly to a DI water supply 3 Water tank spout adapter 1 One adapter is included to neatly allow for 1 4 tubing to quickly connect and dis...

Page 5: ...lure Step 2 Install Dispensing Cap Hand tighten dispensing cap onto water tank Caution DO NOT use any tools over tightening the cap will damage the gasket and cause water leakage Ensure the water tank...

Page 6: ...humidifier is placed on a level surface with no more than 5 tilt Step 2 Turn on Water Valve Ensure feeding tube is fully inserted Switch the water tank tap lever to the ON position to the left for wa...

Page 7: ...the LED will turn green to indicate system has enough water to produce mist Step 3 Connect to Controller Concluding the function test remove the three prong AC line cord from the wall outlet and conn...

Page 8: ...r within an environmental chamber and plug into the rear sensor socket of a Series 5000 controller With the sensor properly installed on the rear of the controller switch the back panel power switch t...

Page 9: ...on 3 While the bi color LED is red the water solenoid is open to let de ionized or distilled water fill the basin to acceptable level CAUTION IF RED LED IS STAYS ON FOR MORE THAN 3 MINUTES UNPLUG THE...

Page 10: ...contaminant build up in the humidifier basin and increase unit lifespan Filter specifications include 99 999 Bacterial Filtration Efficiency BFE 99 999 Viral Filtration Efficiency VFE and a flow resi...

Page 11: ...eing used for more than a week drain the water completely out of the basin Accessory filters have limited lifespan It is recommended to replace the filter regularly to avoid degradation in humidificat...

Page 12: ...to the unit Turn the 5 gallon water tank valve to the ON position to let fresh deionized water from the tank flush out contaminates in the basin through the drain valve for at least 2 minutes Step 3 R...

Page 13: ...urrent humidity Electrical issue resulting in LED not lighting Contact ETS for unit repair Power supply connection not reaching board Board short circuiting Red LOW WATER light stays on Water basin is...

Page 14: ...defective workmanship or material Seller expressly disclaims all other warranties express and or implied including but not limited to those of merchantability and fitness for a particular purpose In n...