4
2. Replace the tank on the humidifier unit.
The water will automatically drain from the tank into the basin and stop
when the basin is full. If the tank is not put on properly, or the humidifier is
not on a level surface, water may continue to flow after the basin is full. If
this happens, remove the top immediately, check for problems and try
again. If the problem persists, contact ETS.
3. Turn on the humidifier POWER switch.
(If used with an ETS Automatic Humidity Controller, see “Humidifying” in
the Controller manual.) The power indicator light will turn on immediately
and the internal blower will begin moving. When the basin is full, the
automatic water level switch will provide power to the ultrasonic
transducer. When this happens, the unit will begin producing a fine mist.
4. Rotate the MIST INTENSITY on the POWER switch.
This control determines how fast the water is converted from a liquid to a
mist. For most applications, set the control at mid-point (II - position).
Rotating the knob 1 place, counter-clockwise (I – position) will produce a
finer mist. When operating at high temperatures or in large enclosures,
set the knob at maximum (II) for best results.
At this point the humidifier should be running properly, blowing a fine mist
into the chamber.
4.2 OPERATION
After following the above set-up procedure, the Model 5462 Ultrasonic
Humidifier should be operating properly. The humidifier should perform
well for a long time if the following precautions are observed:
1. Always run the humidifier directly to the chamber.
Never attempt to combine the humidifier output with another air or gas
source. Any back-pressure in the humidifier system will cause the water
mist to be forced into the electronics compartment, which may damage the
humidifier.