ETS 5462 Operating Manual Download Page 6

 
 

2.  Replace the tank on the humidifier unit.   
 

The water will automatically drain from the tank into the basin and stop 
when the basin is full.  If the tank is not put on properly, or the humidifier is 
not on a level surface, water may continue to flow after the basin is full.  If 
this happens, remove the top immediately, check for problems and try 
again.  If the problem persists, contact ETS. 
  

3.  Turn on the humidifier POWER switch.  
 

(If used with an ETS Automatic Humidity Controller, see “Humidifying” in 
the Controller manual.)  The power indicator light will turn on immediately 
and the internal blower will begin moving.   When the basin is full, the 
automatic water level switch will provide power to the ultrasonic 
transducer.  When this happens, the unit will begin producing a fine mist. 
 

4.  Rotate the MIST INTENSITY on the POWER switch. 
 

This control determines how fast the water is converted from a liquid to a 
mist.  For most applications, set the control at mid-point (II - position).  
Rotating the knob 1 place, counter-clockwise (I – position) will produce a 
finer mist.  When operating at high temperatures or in large enclosures, 
set the knob at maximum (II) for best results. 

 

At this point the humidifier should be running properly, blowing a fine mist 
into the chamber. 

 

4.2     OPERATION  

 

After following the above set-up procedure, the Model 5462 Ultrasonic 
Humidifier should be operating properly.  The humidifier should perform 
well for a long time if the following precautions are observed: 

 
1.  Always run the humidifier directly to the chamber. 

 
Never attempt to combine the humidifier output with another air or gas 
source.  Any back-pressure in the humidifier system will cause the water 
mist to be forced into the electronics compartment, which may damage the 
humidifier. 
 
 
 
 
 

Summary of Contents for 5462

Page 1: ...Ultrasonic Humidification System Model 5462 Model 5472 OPERATING MANUAL 6 30 06 www electrotechsystems com...

Page 2: ...l 5462 03 4 2 Model 5462 Operation 04 5 0 Model 5472 06 5 1 Model 5472 Set Up 07 5 11 Using the Water Tank 07 5 12 Using the Tap Water System 09 5 2 Operation 10 5 2 1 Features 10 5 22 Operating Preca...

Page 3: ...late producing a fine mist An air moving device forces the mist out of the humidifier through a 1 plastic tube and into the enclosure A 4 110 cfm fan is provided it should be run continuously to circu...

Page 4: ...e the chamber If the humidifier is being used with an ETS Chamber the barb is already installed 2 Place the humidifier a minimum of 12 away from the chamber 3 Cut the 1 I D tubing to the appropriate l...

Page 5: ...is available as 115 VAC or 230 VAC 50 60 Hz The voltage must be specified at the time of purchase The circulating fan is supplied at the appropriate voltage 115 VAC or 230 VAC 4 0 MODEL 5462 The Mode...

Page 6: ...ist 4 Rotate the MIST INTENSITY on the POWER switch This control determines how fast the water is converted from a liquid to a mist For most applications set the control at mid point II position Rotat...

Page 7: ...mpartment 3 Never place the humidifier inside the chamber The fans and electronics will become exposed to the wet air and will quickly fail Always operate the humidifier outside the chamber and attach...

Page 8: ...Ultrasonic Humidifiers The Model 5472 has many features not found in the Model 5462 Figure 5 0 1 5 Gallon Water Tank for long term applications Deionizer Filtering Column 10 ft OD Polytubing Quick Co...

Page 9: ...deionizing resin is depleted it will change from dark to light in color When it is light it must be replaced 5 11 Using the Water Tank 1 Attach the water de ionizing column to a faucet The faucet shou...

Page 10: ...the humidifier 5 Attach the provided plastic faucet to the water tank The faucet has a O D tubing quick connect fitting attached to the outlet 6 Make sure the faucet is closed then push the O D tubing...

Page 11: ...urified water Fast flow rate poorly purified water 3 Turn on the water Set the flow rate to the pre determined amount DO NOT OPERATE THE HUMIDIFIER WITH THE FAUCET 100 OPEN High flow rates will cause...

Page 12: ...er in the Model 5472 3 The Model 5472 can be modified to use compressed air or an inert gas as the propulsion source instead of the air pump The 5472 can also be modified for to be a close loop system...

Page 13: ...y 6 0 MAINTENANCE CLEANING Model 5462 5472 1 Always unplug the unit before cleaning 2 Empty the unit of all water Disconnect or remove water tank or tap water source 3 Clean the surface of the transdu...

Page 14: ...l parts of the humidifier when it is to be stored or taken out of service for any period longer than one week Do not seal the water tank in storage Leave the top off to allow the air to completely dry...

Page 15: ...anty becomes null and void should the equipment or any part thereof be abused or modified by the customer of if used in any application other than that for which it was intended This warranty to repla...

Page 16: ...TANTIALLY INCREASES THE CONTENT S SUSCEPTIBILITY TO SHOCK DAMAGE DO NOT PLACE INSTRUMENTS OR ACCESSORIES INSIDE OTHER INSTRUMENTS OR CHAMBERS ELECTRO TECH SYSTEMS INC WILL NOT ASSUME RESPONSIBILITY FO...

Reviews: