![ETS-Lindgren HI-1801 User Manual Download Page 7](http://html.mh-extra.com/html/ets-lindgren/hi-1801/hi-1801_user-manual_2434504007.webp)
Introduction
|
7
1.0 Introduction
The
ETS-Lindgren Model HI-1801 Microwave Survey Meter
is a rugged,
compact, portable instrument that is virtually immune to failure caused by
excessive fields or physical abuse. This instrument is acceptable to the US
Government Center for Devices and Radiological Health (FDA/CDRH) and to all
major microwave oven manufacturers for testing ovens in use and after repair.
This easy to use meter, and the step-by-step instructions in this manual will
enable you to easily and accurately measure leakage from your microwave oven.
Microwave leakage (electromagnetic fields) is detected by an array of eight hot
carrier diodes housed in the large end of the plastic probe. This antenna array
has the unique feature of being able to sum microwave electric fields of any
polarization in a plane perpendicular to the axis of the probe. The antenna lobe
(effective measuring area) is also very broad, making the instrument easy to use
when measuring leakage around an oven door. The spacer cone is designed to
provide 5 cm spacing from the tip of the probe to the center of the array.
The HI-1801 must be used with the spacer cone in place. Using the
instrument without the cone will result in reading errors and may
damage the unit.
Calibration Method
Each meter is calibrated by placing the probe in a CW (continuous wave)
2450 MHz electromagnetic field. The source feeds an anechoic chamber through
a section of waveguide terminating in a slot radiator. Calibration is performed at a
level of 5 mW/cm² and a standard temperature of 75º F.
The calibration is also checked at 1 mW/cm². The accuracy of the field is
determined by comparing with an LCR (local calibration reference) which is
traceable to NIST (National Institute of Standards and Technology) through
FDA / CDRH (Food and Drug Administration / Center for Devices and
Radiological Health).
Summary of Contents for HI-1801
Page 1: ...Model HI 1801 Microwave Survey Meter User Manual ...
Page 4: ...iv This page intentionally left blank ...
Page 6: ...vi This page intentionally left blank ...
Page 12: ...12 Maintenance This page intentionally left blank ...
Page 14: ...14 Specifications This page intentionally left blank ...
Page 22: ...22 Warranty This page intentionally left blank ...
Page 23: ...EC Declaration of Conformity 23 Appendix B EC Declaration of Conformity ...
Page 24: ...24 EC Declaration of Conformity This page intentionally left blank ...