INSTRUMENT CLEANING
- Clean the
instrument regularly with an anti-bacterial
probe wipe to avoid potential bacterial growth.
BATTERY REPLACEMENT
- A battery
symbol indicates that the batteries will need
replacing. The instrument continues to
measure accurately but we recommend that
the batteries be changed as soon as possible.
To replace the batteries, remove the battery
cover using a coin. Remove the batteries by
pulling the battery
retaining clip back
whilst holding the
unit upside down.
Replace both
batteries, positive side
up, replace cover.
ERROR MESSAGES
'Err' will be displayed
if the probe develops a fault. If the error
message remains, contact our service
department for further assistance.
EMC/RFI
- Instrument performance may be
affected if operated within a high frequency
radio field, such as near a mobile phone, or if
subjected to an electrostatic shock.
GUARANTEE
- This instrument carries a
two-year guarantee against defects in either
components or workmanship. During this
period, products that prove to be defective
will, at the discretion of ETI, be either repaired
or replaced without charge. This guarantee
does not apply to probes, where a six-month
period is offered. The product guarantee does
not cover damage caused by fair wear and
tear, abnormal storage conditions, incorrect
use, accidental misuse, abuse, neglect,
misapplication or modification. Full details
of liability are available within ETI's Terms &
Conditions of Sale at etiltd.com/terms. In line
with our policy of continuous development,
we reserve the right to amend our product
specification without prior notice.
INSTRUMENT OPERATION
- The instrument
is switched on by unfolding the probe until
the display shows
the temperature.
The unit is turned
off by folding the
probe back into
the side of the
instrument. The
probe should not
be rotated more
than 180 degrees
as damage will
occur. It is recommended to store the probe,
in the closed position, when not in use.
Apply the probe to the substance to be
measured. For optimum accuracy immerse
at least 70 mm of the probe in the substance.
DO NOT
immerse any of the case parts in
liquid. The probe should be allowed adequate
time to stabilise to the temperature of the
item to be measured.
INSTRUMENT CONFIGURATION
- This
instrument has four user selectable slide
switches (switch bank), adjustable with a
paperclip, and a single push button 'Trim
Adjust' which is located under the battery
cover.
Switch 1
- Not used
Switch 2
- Resolution: To change the
resolution from 0.01 °C (factory default) to 0.1
°C push the switch away from the batteries.
Switch 3
- Auto-off: To disable the auto-off
(factory default – 10 minutes) push the switch
away from the batteries.
Switch 4
- Reading Trim For normal use, we
do not recommend using the 'Trim Adjust'.
For further details visit etiltd.com