ETI REFERENCE THERMAPEN Operating Instructions Download Page 2

INSTRUMENT CLEANING

 - Clean the 

instrument regularly with an anti-bacterial 

probe wipe to avoid potential bacterial growth.

BATTERY REPLACEMENT 

- A battery 

symbol indicates that the batteries will need  

replacing. The instrument continues to 

measure accurately but we recommend that 

the batteries be changed as soon as possible. 

To replace the batteries, remove the battery 

cover using a coin. Remove the batteries by 

pulling the battery 

retaining clip back 

whilst holding the 

unit upside down. 

Replace both 

batteries, positive side 

up, replace cover.

ERROR MESSAGES 

'Err' will be displayed 

if the probe develops a fault. If the error 

message remains, contact our service 

department for further assistance.

EMC/RFI 

- Instrument performance may be 

affected if operated within a high frequency 

radio field, such as near a mobile phone, or if 

subjected to an electrostatic shock.

GUARANTEE

 - This instrument carries a 

two-year guarantee against defects in either 

components or workmanship. During this 

period, products that prove to be defective 

will, at the discretion of ETI, be either repaired 

or replaced without charge. This guarantee 

does not apply to probes, where a six-month 

period is offered. The product guarantee does 

not cover damage caused by fair wear and 

tear, abnormal storage conditions, incorrect 

use, accidental misuse, abuse, neglect, 

misapplication or modification. Full details 

of liability are available within ETI's Terms & 

Conditions of Sale at etiltd.com/terms. In line 

with our policy of continuous development, 

we reserve the right to amend our product 

specification without prior notice.

INSTRUMENT OPERATION

 

- The instrument 

is switched on by unfolding the probe until  

the display shows 

the temperature. 

The unit is turned 

off by folding the 

probe back into 

the side of the 

instrument. The 

probe should not 

be rotated more 

than 180 degrees 

as damage will 

occur. It is recommended to store the probe, 

in the closed position, when not in use.

Apply the probe to the substance to be 

measured. For optimum accuracy immerse 

at least 70 mm of the probe in the substance. 

DO NOT

 immerse any of the case parts in 

liquid. The probe should be allowed adequate 

time to stabilise to the temperature of the 

item to be measured.

INSTRUMENT CONFIGURATION

 - This 

instrument has four user selectable  slide 

switches (switch bank), adjustable with a 

paperclip, and a single push button 'Trim 

Adjust' which is located under the battery 

cover.

Switch 1

 - Not used

Switch 2

 - Resolution:  To change the 

resolution from 0.01 °C (factory default) to 0.1 

°C push the switch away from the batteries.

Switch 3

 - Auto-off: To disable the auto-off 

(factory default – 10 minutes) push the switch 

away from the batteries.

Switch 4

 - Reading Trim For normal use, we 

do not recommend using the 'Trim Adjust'.  

For further details visit etiltd.com

 