![ETI 822-950 User Manual Download Page 4](http://html1.mh-extra.com/html/eti/822-950/822-950_user-manual_2433022004.webp)
Supplied by
Electronic Temperature Instruments Ltd
Easting Close, Worthing, West Sussex BN14 8HQ
01903 202151 · [email protected] · etiltd.com
822-950/03.11.20
SAFETY
!
If using Silicone oil, do not fill to the lip of the bath in one go. The effects of
thermal expansion will cause the volume of oil to expand as the temperature
rises and will cause the bath to overfill, creating a slip and burn hazard. For
this reason we do not recommend leaving the bath unattended when using
Silicone oil.
!
It is recommended the user monitors the depth of the liquid and adjusts
accordingly to ensure this does not occur. An approximate guide for filling is
to fill 3L of oil then heat to the appropriate temperature before topping up
and allowing for further expansion of the new oil.
!
Only test within the front 130mm
3
of the bath, keeping objects free from the
back of the bath where the circulation pump is located.
THERE ARE NO USER SERVICEABLE PARTS INSIDE.
DRAINING THE BATH
Disconnect the power from the calibration bath before draining or have an
empty container with suitable volume (5L) to hand screw the threaded adaptor
into the mouth of the tap and connect the rubber hosing if needed to control
the flow towards a container. Open the tap fully with a 90° turn and allow the
bath to empty.
Once the flow has lessened, twist the handle 90° back, flat with the bath to close.
The tap and the threaded tap/ hose can then be disconnected.
Using a paper towel, dry the remnants of fluid from inside. The bath can now be
filled with a different contact fluid or safely set aside for storage.
GUARANTEE
- This instrument carries a one-year guarantee against defects
in either components or workmanship. During this period, products that
prove to be defective will, at the discretion of ETI, be either repaired or
replaced without charge. The product guarantee does not cover damage
caused by fair wear and tear, abnormal storage conditions, incorrect use,
accidental misuse, abuse, neglect, misapplication or modification. Full details
of liability are available within ETI’s Terms & Conditions of Sale at etiltd.com/
terms. In line with our policy of continuous development, we reserve the right
to amend our product specification without prior notice.