ETI 822-950 User Manual Download Page 4

Supplied by

Electronic Temperature Instruments Ltd

Easting Close, Worthing, West Sussex BN14 8HQ

01903 202151 · [email protected] · etiltd.com

822-950/03.11.20

SAFETY

 

  

!

   If using Silicone oil, do not fill to the lip of the bath in one go. The effects of 

thermal expansion will cause the volume of oil to expand as the temperature 

rises and will cause the bath to overfill, creating a slip and burn hazard. For 

this reason we do not recommend leaving the bath unattended when using 

Silicone oil.

!

  It is recommended the user monitors the depth of the liquid and adjusts 

accordingly to ensure this does not occur. An approximate guide for filling is 

to fill 3L of oil then heat to the appropriate temperature before topping up 

and allowing for further expansion of the new oil.

!

  Only test within the front 130mm

3

 of the bath, keeping objects free from the 

back of the bath where the circulation pump is located.

 

     THERE ARE NO USER SERVICEABLE PARTS INSIDE.

DRAINING THE BATH

 

  

Disconnect the power from the calibration bath before draining or have an 

empty container with suitable volume (5L) to hand screw the threaded adaptor 

into the mouth of the tap and connect the rubber hosing if needed to control 

the flow towards a container. Open the tap fully with a 90° turn and allow the 

bath to empty.

Once the flow has lessened, twist the handle 90° back, flat with the bath to close. 

The tap and the threaded tap/ hose can then be disconnected.

Using a paper towel, dry the remnants of fluid from inside. The bath can now be 

filled with a different contact fluid or safely set aside for storage.

GUARANTEE

 - This instrument carries a one-year guarantee against defects 

in either components or workmanship. During this period, products that 

prove to be defective will, at the discretion of ETI, be either repaired or 

replaced without charge. The product guarantee does not cover damage 

caused by fair wear and tear, abnormal storage conditions, incorrect use, 

accidental misuse, abuse, neglect, misapplication or modification. Full details 

of liability are available within ETI’s Terms & Conditions of Sale at etiltd.com/

terms. In line with our policy of continuous development, we reserve the right 

to amend our product specification without prior notice.

Reviews: