SAFETY INFORMATION

 

Use this instrument only as specified in this manual, otherwise, the protection 

provided by the instrument may be impaired. 

Please note the warnings and cautions listed below.

WARNINGS

BURN HAZARD- DO NOT

 touch the IR target surface of the unit.

The temperature of the IR target surface is the same as the actual temperature 

shown on the display. If the unit is set at 500 °C and the display reads 500 °C, 

the target surface is 500 °C. The top of the instrument case may exhibit extreme 

temperatures for areas close to the IR target surface.

DO NOT

 turn off the unit at temperatures higher than 50 °C. This could create

a hazardous situation. 

DO NOT

 connect and operate this unit without a properly grounded, properly 

polarized power cord.

HIGH VOLTAGE

 is used in the operation of this equipment. Severe injury or death 

may result if personnel fail to observe safety precautions. Before working inside the 

equipment, turn the power off and disconnect the power cord. Overhead clearance 

is required.

DO NOT

 place this instrument under a cabinet or other structure.

DO NOT

 use this unit in conditions other than those listed in the environmental 

conditions section (page 5).

DO NOT

 operate near flammable materials.

Use of this instrument at 

HIGH TEMPERATURES

 

for extended periods of time 

requires caution. Completely unattended high temperature operation is not 

recommended for safety reasons.

CAUTIONS

To avoid possible damage to the instrument, follow these guidelines.

DO NOT

 plug the unit into 220V if the heater switches and fuse holder read

11OV. This action will cause the fuses to blow and may damage the instrument.

DO NOT

 use fluids to clean the target surface.

DO NOT

 change the values of the calibration constants from the factory set values. 

The correct setting of these parameters is important to the safety and proper 

operation of the calibrator.

Summary of Contents for 822-400

Page 1: ...User Manual IR 500 BLACK BODY CALIBRATOR Product Code 822 400...

Page 2: ......

Page 3: ...may result if personnel fail to observe safety precautions Before working inside the equipment turn the power off and disconnect the power cord Overhead clearance is required DO NOT place this instrum...

Page 4: ...tchable between C F which can be used for different temperature unit calibrations The calibrator operates over a range of 50 C to 500 C with a temperature display resolution of 0 1 C 3 GENERAL SPECIFI...

Page 5: ...sively dusty or dirty environment Maintenance and cleaning recommendations can be found in the Maintenance section of this manual The instrument operates safely under the following conditions OPERATIN...

Page 6: ...begin normal operation If the unit fails to operate please check the power connection The heater will start operating to bring the temperature of the calibrator to the set point temperature and the di...

Page 7: ...inlet that plugs into an IEC grounded socket CAUTION Do not plug the unit into 220V if the unit is indicated as 110V This action will cause the fuses to blow and may damage the instrument Fan The fan...

Page 8: ...0verload alarm when the target assembly temperature 4 C higher than the set temperature indicates the heating power off ALM2 0verload alarm when the target assembly temperature 4 C higher than the set...

Page 9: ...se the SET button and press the button the green LED on the second line will indicate the current temperature unit Press DOWN to select F press UP again to select C The chosen temperature unit will fl...

Page 10: ...ed Pressing and holding down the button the temperature will change in 1 increments The green LED display on the front panel will indicate the actual set temperature When the set temperature changes t...

Page 11: ...11...

Page 12: ...12 Supplied by Electronic Temperature Instruments Ltd Easting Close Worthing West Sussex BN14 8HQ 01903 202151 sales etiltd com etiltd com 822 400 24 11 20...

Reviews: