Page 1: ...engineered excellence Etatron GB Ltd Newlin Business Park Exchange Road Lincoln LN6 3AB...

Page 2: ...ress OK button to highlight the word NO followed by the right arrow button until NO changes to the word RESTORE and press the OK button to confirm 4 The word CONFIRM should now be shown with NO below...

Page 3: ...nnel you wish to set the setpoint for Press OK button to change the default setting and after changing press using left right arrows press OK to confirm Using the left right arrows for the setup of Wo...

Page 4: ...on to change the default setting and after changing press using left right arrows press OK to confirm Using the left right arrows for the setup of Working Mode acid or alkaline Mode on off or timed pu...

Page 5: ...r channels if you have eSelect M3 Press the OK button to begin ZERO calibration Place clean electrode in buffer solution 7 and wait 2 3 minutes Press OK to highlight the pH reading then using the left...

Page 6: ...ng a blanking plug or by short circuiting the BNC connector Press OK to highlight the mV reading then using the left and right arrow buttons adjust the display to match 0 mV and confirm by pressing th...

Page 7: ...t M From main screen select OK and using the right arrow Keep scrolling until screen on the right is shown Press the OK button to begin calibration Take a sample of water from the flow cell and Above...

Page 8: ...vice Dosing setpoint is the recommended option Confirm with OK and then press the right arrow to move to next screen Setpoint Value should now be shown so adjust to match the current setpoint for the...

Page 9: ...mitted time to meet the required setpoint If the set time is reached the instrument outputs will be shut and the relay output can then be set to trigger an alarm if required Next screen is Max Dosing...