3   

8.  At this point the device is ready for use and will start registering the CO

2

 

removal. The equipment at will show you the main screen. 

                                     

 

9. 

Settings during Treatment 

a) 

GAS Flow

 

Selecting the 

 and 

 buttons you can choose the desired air flow. You 

can range between 4 to 15 L/ min or you can reset the flow.

 

By clicking on the 

 button, you can confirm the value and the 

 button 

you can exit the window and selecting 

 you can reset the flow.

 

                                    

 

 

b) 

CO

2

 Removal

 

Selecting  the 

 and 

 you  choose  the  desired  value  of  VCO2  to 

extract. By clicking on the 

 you confirm the value and the 

 button you 

exit the dialog and selecting the 

 button you will disable this feature. 

 

Operating Manual ProLUNG Meter   

 

 

 

 

 

Pag. 20/39 

Rev4 – 2012-12-31 
 

The equipment at this point will present the main screen shown in the figure below: 
 

 

Figure 16 

Use the touch screen to select one of the following functions: 
Air Flow 
CO2 Removed 
Air Temperature 
Treatment Data  
 

 Figure 17 

Selecting the + and - buttons you can choose the desired air flow. The allowable 
values are from 4 to 15 L/ min or you can reset the flow. 
By clicking on the 

!

 button, you can confirm the value and the 

"

 button you can 

exit the window and selecting Off 

#

 you can reset the flow. 

 

Operating Manual ProLUNG Meter                

 

 

 

 

Pag. 21/39 

Rev4 – 2012-12-31 
 

 

Figure 18 

Selecting the + and - you choose the desired value of VCO

2

 to extract. 

By  clicking  on  the 

!

  you  confirm  the  value  and  the 

"

  button  you  exit  the  dialog 

and selecting the Off button 

#

 disables this feature. 

 

 Figure 19 

 
Setting  using  the  +  and  -  you  choose  the  desired  value  of  the  air  temperature 
supplied to the ProLUNG cartridge. By clicking on the 

!

 you confirm the value and 

the 

"

 button exits the window and selecting the 

#

 Off button turns off the heater. 

The  graph  of  VCO

2

  can  be  changed  by  selecting  the  scale  of  the  individual 

axes. 

 

Operating Manual ProLUNG Meter                

 

 

 

 

Pag. 21/39 

Rev4 – 2012-12-31 
 

 

Figure 18 

Selecting the + and - you choose the desired value of VCO

2

 to extract. 

By  clicking  on  the 

!

  you  confirm  the  value  and  the 

"

  button  you  exit  the  dialog 

and selecting the Off button 

#

 disables this feature. 

 

 Figure 19 

 
Setting  using  the  +  and  -  you  choose  the  desired  value  of  the  air  temperature 
supplied to the ProLUNG cartridge. By clicking on the 

!

 you confirm the value and 

the 

"

 button exits the window and selecting the 

#

 Off button turns off the heater. 

The  graph  of  VCO

2

  can  be  changed  by  selecting  the  scale  of  the  individual 

axes. 

 

Operating Manual ProLUNG Meter                

 

 

 

 

Pag. 21/39 

Rev4 – 2012-12-31 
 

 

Figure 18 

Selecting the + and - you choose the desired value of VCO

2

 to extract. 

By  clicking  on  the 

!

  you  confirm  the  value  and  the 

"

  button  you  exit  the  dialog 

and selecting the Off button 

#

 disables this feature. 

 

 Figure 19 

 
Setting  using  the  +  and  -  you  choose  the  desired  value  of  the  air  temperature 
supplied to the ProLUNG cartridge. By clicking on the 

!

 you confirm the value and 

the 

"

 button exits the window and selecting the 

#

 Off button turns off the heater. 

The  graph  of  VCO

2

  can  be  changed  by  selecting  the  scale  of  the  individual 

axes. 

 

Operating Manual ProLUNG Meter                

 

 

 

 

Pag. 21/39 

Rev4 – 2012-12-31 
 

 

Figure 18 

Selecting the + and - you choose the desired value of VCO

2

 to extract. 

By  clicking  on  the 

!

  you  confirm  the  value  and  the 

"

  button  you  exit  the  dialog 

and selecting the Off button 

#

 disables this feature. 

 

 Figure 19 

 
Setting  using  the  +  and  -  you  choose  the  desired  value  of  the  air  temperature 
supplied to the ProLUNG cartridge. By clicking on the 

!

 you confirm the value and 

the 

"

 button exits the window and selecting the 

#

 Off button turns off the heater. 

The  graph  of  VCO

2

  can  be  changed  by  selecting  the  scale  of  the  individual 

axes. 

 

Operating Manual ProLUNG Meter                

 

 

 

 

Pag. 21/39 

Rev4 – 2012-12-31 
 

 

Figure 18 

Selecting the + and - you choose the desired value of VCO

2

 to extract. 

By  clicking  on  the 

!

  you  confirm  the  value  and  the 

"

  button  you  exit  the  dialog 

and selecting the Off button 

#

 disables this feature. 

 

 Figure 19 

 
Setting  using  the  +  and  -  you  choose  the  desired  value  of  the  air  temperature 
supplied to the ProLUNG cartridge. By clicking on the 

!

 you confirm the value and 

the 

"

 button exits the window and selecting the 

#

 Off button turns off the heater. 

The  graph  of  VCO

2

  can  be  changed  by  selecting  the  scale  of  the  individual 

axes. 

 

Operating Manual ProLUNG Meter   

 

 

 

 

 

Pag. 20/39 

Rev4 – 2012-12-31 
 

The equipment at this point will present the main screen shown in the figure below: 
 

 

Figure 16 

Use the touch screen to select one of the following functions: 
Air Flow 
CO2 Removed 
Air Temperature 
Treatment Data  
 

 Figure 17 

Selecting the + and - buttons you can choose the desired air flow. The allowable 
values are from 4 to 15 L/ min or you can reset the flow. 
By clicking on the 

!

 button, you can confirm the value and the 

"

 button you can 

exit the window and selecting Off 

#

 you can reset the flow. 

 

Operating Manual ProLUNG Meter                

 

 

 

 

Pag. 21/39 

Rev4 – 2012-12-31 
 

 

Figure 18 

Selecting the + and - you choose the desired value of VCO

2

 to extract. 

By  clicking  on  the 

!

  you  confirm  the  value  and  the 

"

  button  you  exit  the  dialog 

and selecting the Off button 

#

 disables this feature. 

 

 Figure 19 

 
Setting  using  the  +  and  -  you  choose  the  desired  value  of  the  air  temperature 
supplied to the ProLUNG cartridge. By clicking on the 

!

 you confirm the value and 

the 

"

 button exits the window and selecting the 

#

 Off button turns off the heater. 

The  graph  of  VCO

2

  can  be  changed  by  selecting  the  scale  of  the  individual 

axes. 

 

Operating Manual ProLUNG Meter                

 

 

 

 

Pag. 21/39 

Rev4 – 2012-12-31 
 

 

Figure 18 

Selecting the + and - you choose the desired value of VCO

2

 to extract. 

By  clicking  on  the 

!

  you  confirm  the  value  and  the 

"

  button  you  exit  the  dialog 

and selecting the Off button 

#

 disables this feature. 

 

 Figure 19 

 
Setting  using  the  +  and  -  you  choose  the  desired  value  of  the  air  temperature 
supplied to the ProLUNG cartridge. By clicking on the 

!

 you confirm the value and 

the 

"

 button exits the window and selecting the 

#

 Off button turns off the heater. 

The  graph  of  VCO

2

  can  be  changed  by  selecting  the  scale  of  the  individual 

axes. 

 

Operating Manual ProLUNG Meter                

 

 

 

 

Pag. 21/39 

Rev4 – 2012-12-31 
 

 

Figure 18 

Selecting the + and - you choose the desired value of VCO

2

 to extract. 

By  clicking  on  the 

!

  you  confirm  the  value  and  the 

"

  button  you  exit  the  dialog 

and selecting the Off button 

#

 disables this feature. 

 

 Figure 19 

 
Setting  using  the  +  and  -  you  choose  the  desired  value  of  the  air  temperature 
supplied to the ProLUNG cartridge. By clicking on the 

!

 you confirm the value and 

the 

"

 button exits the window and selecting the 

#

 Off button turns off the heater. 

The  graph  of  VCO

2

  can  be  changed  by  selecting  the  scale  of  the  individual 

axes. 

 

Operating Manual ProLUNG Meter                

 

 

 

 

Pag. 21/39 

Rev4 – 2012-12-31 
 

 

Figure 18 

Selecting the + and - you choose the desired value of VCO

2

 to extract. 

By  clicking  on  the 

!

  you  confirm  the  value  and  the 

"

  button  you  exit  the  dialog 

and selecting the Off button 

#

 disables this feature. 

 

 Figure 19 

 
Setting  using  the  +  and  -  you  choose  the  desired  value  of  the  air  temperature 
supplied to the ProLUNG cartridge. By clicking on the 

!

 you confirm the value and 

the 

"

 button exits the window and selecting the 

#

 Off button turns off the heater. 

The  graph  of  VCO

2

  can  be  changed  by  selecting  the  scale  of  the  individual 

axes. 

 

Operating Manual ProLUNG Meter                

 

 

 

 

Pag. 21/39 

Rev4 – 2012-12-31 
 

 

Figure 18 

Selecting the + and - you choose the desired value of VCO

2

 to extract. 

By  clicking  on  the 

!

  you  confirm  the  value  and  the 

"

  button  you  exit  the  dialog 

and selecting the Off button 

#

 disables this feature. 

 

 Figure 19 

 
Setting  using  the  +  and  -  you  choose  the  desired  value  of  the  air  temperature 
supplied to the ProLUNG cartridge. By clicking on the 

!

 you confirm the value and 

the 

"

 button exits the window and selecting the 

#

 Off button turns off the heater. 

The  graph  of  VCO

2

  can  be  changed  by  selecting  the  scale  of  the  individual 

axes. 

Reviews: