D
IVISION
D
ESCRIPTION
T
EST
METHOD
Schober's test -
(Decussation circle)
The test is for
measuring
vertical/horizontal
phoria of right or
left eye by using
Red/Green filter.
1. Show Schober's test type to a patient.
2. Let a patient wear Red/Green glasses (right eye = red filter,
left eye = green filter). The patient sees «
» ”mark in a right
eye and «
» mark in a left eye on Schober's test type. The
binocular condition would be ideal, if the patient could see
«
» mark.
3. Judge a patient's binocular function as below, while asking
him how the type is seen.
Patient's visual
status
Fixation
disparity
Method
Esophoria
ⓐ
Add B.O prism until being
no deviation.
Exophoria
ⓑ
Add B.I prism until being
no deviation.
Right eye's
ⓒ
hypophoria / Left
eye's hyperphoria
Add B.U in a right eye and
B.D in a left eye until
being no deviation.
Right eye's
ⓓ
hyperphoria / Left
eye's hypophoria
Add B.U in a left eye and
B.D prism in a right eye
until being no deviation.
U
SER
MANUAL
> VII. T
EST
METHOD
27
CSPOLA600 / CSPOLA600C - Polarized visual chart system > V5 - 03-2018
Summary of Contents for CS pola 600
Page 1: ...www essilor instruments com User manual...
Page 3: ...I INTRODUCTION...
Page 6: ...II SAFETY INFORMATION...
Page 11: ...III FEATURES...
Page 13: ...IV NOTES FOR USING THE INSTRUMENT...
Page 15: ...V CONFIGURATIONS...
Page 17: ...VI INSTALLATION...
Page 19: ...VII TEST METHOD...
Page 24: ...VIII REMOTE CONTROL FUNCTION...
Page 27: ...IX HOW TO RECHARGE A REMOTE CONTROL...
Page 29: ...X UPGRADING A CSPOLA600 CSPOLA600C SYSTEM...
Page 31: ...XI SPECIFICATIONS...
Page 33: ...XII COMPONENT LIST...
Page 35: ...XIII SERVICE INFORMATION...