4. C
LASSIFICATION
Classification
This device is compliant with
marking.
Date of first marking: September 2018
Class I medical device
The expected life of the device and its components is 4 years.
a. Technical features
Technical data
Value
Supply voltage
100-240 VAC
Rated frequency
50/60 Hz
Internal voltage
24 VDC - 12 VDC - 5 VDC
Maximum absorbed power
52 VA
Maximum absorbed current
400 mA
Safety class
I
Applied part
Type BF
Dimensions
340 x 430 x 460 mm
Weight
14.7 kg
Sensor resolution
2.1 mpixels
b. Environmental conditions
Temperature
Humidity
Atmospheric pressure
Use
[10°C ; 40°C]
Maximum 90 % without condensation
[800hPa ; 1060hPa]
Storage
[0°C ; 60°C]
Maximum 90 % without condensation
[700hPa ; 1060hPa]
Transportation
[-10°C ; 60°C]
Maximum 90 % without condensation
[500hPa ; 1060hPa]
Lighting: for a more successful result, it is recommended to perform the exam in a dimly lit area.
The device does not meet the temperature requirements of ISO 15004-1 for storage. Do not store
the device in conditions where the temperature may exceed 60°C or drop lower than 0°C.
Retina800 - Fundus camera > V2 - 06-2019
10
U
SER
MANUAL
> II. S
AFETY
Summary of Contents for Retina 800
Page 1: ...User manual...
Page 4: ...I INTRODUCTION...
Page 6: ...II SAFETY...
Page 14: ...III SUPPLY PACKAGE...
Page 18: ...IV INSTALLATION OF THE DEVICE...
Page 30: ...V USE OF THE DEVICE...
Page 58: ...VI SETTINGS...
Page 101: ...VII MAINTENANCE...
Page 103: ...VIII TROUBLESHOOTING...
Page 106: ...IX ASSISTANCE NOTIFICATION ALERT AND LABEL...
Page 108: ...X GUIDANCE AND MANUFACTURER S DECLARATION...
Page 112: ...XI QR CODE...