Read the instructions for use of the software.
The PC must be compliant to the norm IEC 60950-1 Information technology equipment - Safety -
Part 1: General requirements.
If the PC is installed in the patient area it is necessary to install an isolation electrical supply
compliant with the directive IEC 60601-1 and 3.1 - Medical electrical equipment - Part 1: General
requirements for basic safety and essential performance.
It is possible to connect other accessories to the PC (printer, modem, scanner, etc) through the
analogical or digital interfaces.
The accessories (printer, modem, scanner, etc) must be installed outside the patient area.
The accessories must be compliant to the norm IEC 60950-1 Information technology equipment -
Safety - Part 1: General requirements.
If the accessories are installed in the patient area it is necessary to install an isolation electrical
supply compliant with the directive IEC 60601-1 and 3.1 - Medical electrical equipment - Part 1:
General requirements for basic safety and essential performance.
Patient are is the volume in which the patient can come into contact (intentional or not, direct or
through the operators) with medical electrical an other devices.
4. M
EDICAL
DEVICES
CLASSIFICATION
Technical data
Value
Classification in compliance with the attached IX to the
Directive 93/42/EEC and successive modifications
Class IIa
Retina550 - Fundus camera > V1 - 06-2017
14
U
SER
MANUAL
> II. S
AFETY
Summary of Contents for Retina 550
Page 1: ...www essilor instruments com User manual...
Page 3: ...I INTRODUCTION...
Page 8: ...II SAFETY...
Page 16: ...III DEVICE DESCRIPTION...
Page 20: ...IV DEVICE USE...
Page 29: ...V ORDINARY MAINTENANCE...
Page 31: ...Essilor International 147 rue de Paris 94220 Charenton le Pont France www essilor com...