Division
Description
Test method
Phoria test -
(Fixation point of
cross phoria)
This test is for
measuring
vertical/horizontal
fixation disparity of
right or left eye by
using Red/Green
filter.
1. Show cross phoria test type to a patient.
2. Let a patient wear Red/Green glasses (right eye = red filter, left
eye = green filter). The patient sees «
» mark in a right eye and
«
» mark in a left eye on cross phoria test type. The binocular
condition would be ideal, if the patient could see «
» mark.
3. Judge a patient's binocular function as below, while asking him
how the type is seen.
Patient's
visual status
Fixation disparity
Method
Esodisparity
ⓐ
Add B.O prism until being no
deviation.
Exodisparity
ⓑ
Add B.I prism until being no
deviation.
Left eye's
ⓒ
hyperdisparity
Right eye's
hypodisparity
Add B.U in a right eye or B.D
in a left eye until being no
deviation.
Right eye's
ⓓ
hyperdisparity
Left eye's
hypodisparity
Add B.U in a left eye or B.D
prism in a right eye until
being no deviation.
CSPOLA600 - Polarized visual chart system > V5 - 03-2018
42
U
SER
MANUAL
> VII. T
EST
METHOD
Summary of Contents for CS POLA 600
Page 1: ...www essilor instruments com User manual...
Page 4: ...I INTRODUCTION...
Page 8: ...II SAFETY INFORMATION...
Page 18: ...III FEATURES...
Page 20: ...IV NOTES FOR USING THE INSTRUMENT...
Page 22: ...V CONFIGURATIONS...
Page 25: ...VI INSTALLATION...
Page 37: ...VII TEST METHOD...
Page 43: ...VIII REMOTE CONTROL FUNCTION...
Page 46: ...IX HOW TO RECHARGE A REMOTE CONTROL...
Page 48: ...X UPGRADING A CSPOLA600 SYSTEM...
Page 50: ...XI MAINTENANCE OF UNIT...
Page 52: ...XII TROUBLESHOOTING GUIDE...
Page 54: ...XIII ENVIRONMENTAL CONDITIONS...
Page 56: ...XIV SPECIFICATIONS...
Page 58: ...XV COMPONENT LIST...
Page 60: ...XVI SERVICE INFORMATION...
Page 62: ...Essilor International 147 rue de Paris 94220 Charenton le Pont France www essilor com...