16
ess MaxCharge™ spraygun Owner’s Manual
Troubleshooting Guide
when you encounter the problems listed below, use the suggested trouble-shooting methods. If you
cannot solve the problem or have a problem with the spraygun that is not addressed in this manual,
contact ess at (706) 769-0025.
air pressure of spray
· Clogged liquid Filter
· Clean the liquid Filter assembly
appears
low
(see
the
Liquid Filter Assembly
section of this manual)
· liquid fittings are loose
· Make sure the liquid Filter assembly is installed correctly.;
· air fittings are loose
· Make sure that the air Filter is installed correctly.
or damaged
Inspect for loose hoses or failed air lines –
spray fittings with soapy water – tighten ones that bubble
replace
if
necessary
· Dirty Trigger
· Clean the Trigger
(see
the
To Clean the Trigger
section of this manual)
no spray from nozzle or
· Debris in the nozzle
· Clean nozzle according to instructions
the spray from nozzle is
erratic or spits
· spray is freezing due to
·Make sure that water temperature is at least 50° (10°C)
evaporative
cooling
· liquid filters are clogged
· Clean the liquid Filter assembly
(see
the
Liquid Filter Assembly
section of this manual)
· loose liquid or air fitting
· Inspect hose quick connects at case and at spraygun
leader. Make sure that all liquid fittings and air fittings are
properly
seated.
· Dirty Trigger
· Clean the Trigger
(see
the
To Clean the Trigger
section of this manual)
· Overtightened nozzle cover
· loosen cover. It should only be finger-tight
Charging indicator (leD)
· Batteries are exhausted
recharge batteries
blinks or goes is out
If problem persists, replace battery pack
during operation
· Dirty nozzle
· Clean nozzle according to instructions
Symptom
Possible Problem(s)
Corrective Action
A note about operating temperatures
The MaxCharge nozzle should always be operated at temperatures above 10° Celsius (50° Fahrenheit).
when the ambient temperature is colder than this, the evaporative cooling caused as the spray is
atomized will freeze the nozzle opening.
nozzle freeze-up can also occur when the liquid to be sprayed is colder than 10° C (50° F).