![Esogetics synapsis wave 2 Instruction Manual Download Page 37](http://html1.mh-extra.com/html/esogetics/synapsis-wave-2/synapsis-wave-2_instruction-manual_2427244037.webp)
Guidance and manufacturer’s declaration - electromagnetic immunity
The synapsis wave 2 is intended for use in the electromagnetic environment
specified below. The customer or the user of the synapsis wave 2 should
assure that it is used in such an environment(wave healthcare environment
– excluding outdoors, vehicles and public places).
Immunity test
IEC 60601 test level
Compliance
level
Electromagnetic
environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV
± 15 kV
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material,
the relative
humidity should be
at least 30 %.
magnetic field
Power
frequency
(50/60) Hz
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency
magnetic fields
should be at levels
characteristics of a
typical location in a
typical commercial
or hospital
environment.
Conducted RF
IEC 61000-4-6
V1 = 6 Vrms
150 kHz – 80 MHz
V1 =
6 Vrms
Radiated RF
IEC 61000-4-3
E1 = 10 V/m
for wave healthcare
environment
80 MHz – 2,7 GHz
10 V/m
Field strengths from
fixed RF
transmitters, as
determined by an
electromagnetic site
survey, should be
less than the
compliance level in
each frequency
range. Interference
may occur in the
vicinity of
equipment
37
Summary of Contents for synapsis wave 2
Page 1: ...synapsis wave 2 BEDIENUNGSANLEITUNG INSTRUCTION MANUAL ISTRUZIONI PER L USO...
Page 2: ...2...
Page 14: ...14...
Page 15: ...15 Anhang 1 EMV Tabellen...
Page 18: ...18...
Page 20: ...20...
Page 21: ...ENGLISH...
Page 22: ......
Page 34: ...34...
Page 35: ...35 Appendix 1 EMV Tables...
Page 38: ...38...
Page 40: ......
Page 41: ...41 ITALIANO...
Page 42: ...42...
Page 54: ...54...
Page 55: ...55 Allegato 1 Tabelle EMV...
Page 58: ...58...