5.1.3 Science templates
Slit observations
5.1.3.1.1 Staring
Table 16: User defined and fixed parameters for
SHOOT_slt_obs_Stare.
SHOOT_slt_obs_Stare
Keyword
Range
Default
Label in P2PP
Free parameters
INS.OPTI3.NAME
see
Table 4
NODEFAULT
UVB Slit slide
INS.OPTI4.NAME
see
Table 7
NODEFAULT
VIS Slit slide
INS.OPTI5.NAME
see
Table 8
NODEFAULT
NIR Slit slide
DET1.WIN1.UIT1
0..3600
NODEFAULT
UVB Exposure Time (s)
DET1.READ.CLKDESCR
see
Table 5
100k/1pt/hg
UVB read-out mode
DET2.WIN1.UIT1
0..3600
NODEFAULT
VIS Exposure Time (s)
DET2.READ.CLKDESCR
see
Table 5
100k/1pt/hg
VIS read-out mode
DET3.DIT
0.88..3600
NODEFAULT
NIR Detector Integration
Time (s)
DET3.NDIT
1..9999
1
number of DITs
SEQ.NEXP.UVB
0..999
1
UVB number of exposure
SEQ.NEXP.VIS
0..999
1
VIS number of exposure
SEQ.NEXP.NIR
0..999
1
NIR number of exposure
SEQ.AGSNAPSHOT
T,F
F
Take an acquisition image
before science exposures?
Fixed Values
INS.MODE
SLITSPEC,
IFUSPEC
SLITSPEC
Instrument Mode