Oxygen detectors – URS20SE 

16/06/2017 

Gas detection systems for industrial environments 

4/8

 

 

Electrical Installation 

 
 

 
 
 
 
 
 
 

Terminal board and 
electrical connections 
 
 
 
 
 
 
 
 
 

 
 

Cabling:

 

 
 
 
 

Configuration: 

 
 
 

4…20mA Output 
reference selection: 
 
 
 
 
 
 
4...20mA signal 
operating mode 
configuration: 
 
 
 
Setting threshold limit 
values: 
 
 

CAUTION:  

 Make the area safe and make sure that the device power supply is 

off before cabling and configuration operations. 

Install the sensor in compliance with EN 60079-14. 
ATEX  certified  1“  NPT  cable  glands  are  used  for  cable  entry,  in  compliance  with 
standards EN 60079-0 and EN 60079-1 (Ex d protection mode). 
Ground the sensor using the internal grounding system. 

 

Relay card  
(on connector 
CN3) 

Terminal Board TB1 

12…24Vdc 

4…20mA 

 

JP2 jumper circuit                                                                                    Sensor card 

 

S1 DIP switch 

 

CN4 connector for 
Handheld terminal 

 

 
Depending  on  the  connecting  distance,  use  at  least  3-core  cable,  min.  diameter 
0.75mm

2

up to 100m, 1mm

2

 up to 200m, 1.5mm

2

 up to 500m. 

Use shielded cable where there is a risk of electromagnetic interference.  
If a relay card is used, use multi-core cable suitable for the number of connections. 
Make sure that the cable sheath is no larger than the diameter of the cable gland. 

 

Default set

tings of the sensor are shown in the “Technical Specifications” chapter. 

In  order  to  change  the  default  settings,  switch  off  the  power  supply,  input  the  new 
settings at the JP2 jumper circuit or at the  S1 DIP switch as shown in the diagram, 
then power-up again; in particular: 

 

The  default  setting  for  the  4-20mA  signal  is  the  negative  power  signal.  Output 
reference selection should be made by JP2 triple of jumpers; to change this setting, 
operator has to move 

JP2

 jumpers as shown in the figure: 

Caution:  if  the  default  settings  are  changed, 
the  connections  on  the 

TB1

  terminal  board 

will be inverted. 

     Ref. at  

-

  (default)            Ref. at    

 

 

To  set  the  operating  mode  of  the  4...20mA  signal,  it  is  necessary  to  use  the  5th 
selector of the DIP switch at 

S1

; in particular: 

 
 

 

                                            Proportional                        Threshold mode 

 

To  set  the  threshold  limit  values  of  the  optional  relay  card,  or  of  the  threshold 
operating mode of the 4...20mA signal, it is necessary to use the last selector of the 
DIP switch at 

S1

 (ignore the first four selectors); in particular, the thresholds, given in 

percentage of O

2

, will be: 

 

 

 
(S1.8 = OFF, default): 

Pre-alarm 

1

st

 threshold alarm 

2

nd

 threshold alarm 

O

2

 excess

 

 

23% O

2

 

25% O

2

 

27% O

2

 

O

2

 deficiency

 

 

19% O

2

 

17% O

2

 

15% O

2

 

 
(S1.8 = ON): 

Pre-alarm 

1

st

 threshold alarm 

2

nd

 threshold alarm 

 
 

22% O

2

 

23% O

2

 

24% O

2

 

 
 

20% O

2

 

19% O

2

 

18% O

2

 

1   2   3   4   5   6   7   8 

ON 

1   2   3   4   5   6   7   8 

ON 

Connection slot 
for relay card 

Summary of Contents for Sensigas URS20SE

Page 1: ...entres and more in general where oxygen is stored Oxygen deficiency is an indirect measurement of the presence of other explosive or asphyxiant gases that deplete the oxygen in the air URS20SE sensors...

Page 2: ...default Consumption mode applications at 1 or 2 thresholds 4 20mA Output reference selection 4mA 0 O2 20mA 30 O2 0mA no alarm 10mA 1st threshold alarm 20mA 2nd threshold alarm by jumper selectable pol...

Page 3: ...n this case the detectors must be located at the breathing height of the occupants in the premises Take into consideration the following specific installation guidelines as well as the above instructi...

Page 4: ...iameter of the cable gland Default settings of the sensor are shown in the Technical Specifications chapter In order to change the default settings switch off the power supply input the new settings a...

Page 5: ...ions below Step 1 Insert the connection slot provided with the relay card into the control card making sure the flexible tab is towards the main terminal board Find CN3 connector CN3 connector Connect...

Page 6: ...factory calibrated so they normally do not require any other calibration once installed Still after installation an operational check of the sensors is recommended The detector will enter a 2 minute w...

Page 7: ...rned to the supplier installer who in turn will send it back to the manufacturer Sensors may need to be re calibrated using the TUL40 calibration kit and the TUS40 20 handheld terminal which must be c...

Page 8: ...industry 2 Equipment category 2 for use in Zone 1 G Equipment intended for use in explosive gas atmosphere caused by mixture of air and gas vapours flammable mist Ex d IIC T6 Type of protection compl...