ESD E-003 Instruction Manual Download Page 9

CAUTION

ELECTRICAL SHOCK HAZARD

Do not insert objects through intake or outlet grille.
Do not try to verify operation of unit by drawing a spark from an ion
emitter point. The design of the balancing circuit makes the

park

test

inconclusive. Sustained grounding of ion emitters may

damage the balancing circuit.

6.4 Ion Balance Check

Use a charged plate monitor to test unit for ion balance of +/- 10V.
Do not try to determine ion balance by holding a static meter in the
ionized air stream. This will result in a meaningless reading.

6.5 Calibration

Ion output of the E-003 unit is inherently balanced by design.

As a result, there are no calibration adjustment. After checking ion
balance as outlined in Section 5.4,if an unbalance or offset voltage
exists in excess of 0

±

10 volts, contact Customer Service Center.

6.6 Task Light Replacement

If either task light doesn

t turn on, dims appreciably or

flickers, the compact fluorescent lamp should be replaced as soon as
possible. Access to lamp is provided by removing the front panel of
the device. Disconnect the device by unplugging line cord. Remove
appropriate lamp access panel with a screwdriver and remove light
bulb. Install a new light bulb, replace lamp access panel and plug
back in the device.

ESD 

Conductive 

Co.,Ltd

Summary of Contents for E-003

Page 1: ...INSTRUCTION E 003 IONIZING AIR BLOWER E S D C o n d u c t i v e C o L t d...

Page 2: ...unit designed to provide total coverage for electrostatic problems Please read the instruction manual completely before installing or operating the device This device is equipped with a 3 wires ground...

Page 3: ...2 STRUCTURE E S D C o n d u c t i v e C o L t d...

Page 4: ...ximum air volume According to ESD STM3 1 2000 Ionization Air Coverage 30cm 150cm Working Temperature 0 50 Fan Speed 250 500FPM 30cm Continuously adjustable Noise 50 69dB 60cm Indicator Light Illuminat...

Page 5: ...ng tabs Chains and mounting hardware not supplied must have a minimum safe working load of 15kg 4 2 Electrical The E 003 requires 220VAC 50Hz or 110VAC 60Hz for proper operation The device must be gro...

Page 6: ...work surfaces tools materials and components from developing a static charge Charged items introduced into a work area will be neutralized and will remain neutral while keeping in the ionized air str...

Page 7: ...leaned with a soft brush or vacuum 6 3 Ion Output Check Use a charged plate monitor to test the unit for ion output Discharge times should be measured and checked against the following Ion Output tabl...

Page 8: ...stic from ionized air stream and measure static charge The plastic should be neutralized If no instrumentation is available ion output may be verified with the following procedure 1 Tear off about a 3...

Page 9: ...tion Ion output of the E 003 unit is inherently balanced by design As a result there are no calibration adjustment After checking ion balance as outlined in Section 5 4 if an unbalance or offset volta...

Reviews: