CAUTION
ELECTRICAL SHOCK HAZARD
Do not insert objects through intake or outlet grille.
Do not try to verify operation of unit by drawing a spark from an ion
emitter point. The design of the balancing circuit makes the
“
park
test
”
inconclusive. Sustained grounding of ion emitters may
damage the balancing circuit.
6.4 Ion Balance Check
Use a charged plate monitor to test unit for ion balance of +/- 10V.
Do not try to determine ion balance by holding a static meter in the
ionized air stream. This will result in a meaningless reading.
6.5 Calibration
Ion output of the E-003 unit is inherently balanced by design.
As a result, there are no calibration adjustment. After checking ion
balance as outlined in Section 5.4,if an unbalance or offset voltage
exists in excess of 0
±
10 volts, contact Customer Service Center.
6.6 Task Light Replacement
If either task light doesn
’
t turn on, dims appreciably or
flickers, the compact fluorescent lamp should be replaced as soon as
possible. Access to lamp is provided by removing the front panel of
the device. Disconnect the device by unplugging line cord. Remove
appropriate lamp access panel with a screwdriver and remove light
bulb. Install a new light bulb, replace lamp access panel and plug
back in the device.
ESD
Conductive