Common Diagnostic Techniques
45
2.
The probe becomes part of the circuit during near-field measurements.
There is capacitance and inductance between the circuit being
measured and the probe with the associated cabling. The probe will
re-radiate the received field, altering the field being measured.
However, technical imprecision does not necessarily eliminate a method.
Sometimes an attenuation of the field strength in the near-field will translate into
an attenuation of the far-field reading. As long as a linear relationship is not
expected, there can be real benefit from near-field probing. Generally, a
reduction of the non-radiating field will also mean that the radiating field has been
reduced.
Summary of Contents for ETS-Lindgren 7405
Page 1: ...Model 7405 Near Field Probe Set User Manual...
Page 2: ...ii www ets lindgren com...
Page 12: ...12 Maintenance www ets lindgren com This page intentionally left blank...
Page 18: ...18 Operation www ets lindgren com This page intentionally left blank...
Page 20: ...20 Typical Performance Factors www ets lindgren com Magnetic H Field Probes 901 6 CM LOOP...
Page 21: ...www ets lindgren com Typical Performance Factors 21 902 3 CM LOOP...
Page 22: ...22 Typical Performance Factors www ets lindgren com 903 1 CM LOOP...
Page 23: ...www ets lindgren com Typical Performance Factors 23 Electric E Field Probes 904 BALL PROBE...
Page 24: ...24 Typical Performance Factors www ets lindgren com 905 STUB PROBE...
Page 25: ...www ets lindgren com Typical Performance Factors 25 Preamplifier Gain 0 01 MHZ 3 GHZ...
Page 26: ...26 Typical Performance Factors www ets lindgren com This page intentionally left blank...
Page 49: ...www ets lindgren com EC Declaration of Conformity 49 Appendix A EC Declaration of Conformity...