MIRI
®
II-12 Multiroom IVF incubators User Manual Rev. 4.0
80
Please follow these safety precautions when changing regulators:
•
Always use original pressure regulators (contact Esco Medical or your local distribu-
tor for more details or ordering).
•
Change the regulators within 4 years from the date of installation.
•
Failure to change the regulators may cause breakdowns.
•
Warranty void if wrong/ non-original regulators are used.
Please refer to the service manual for replacement instructions.
37.13 Internal 0.2µ filter for CO
2
gas
The round-shape in line 0.2µ HEPA filter for CO
2
gas removes any particles found in the
incoming CO
2
gas. Failure to use the HEPA filter may cause damage to the high precision flow
sensor, calculate the amount of CO
2
gas entering the system, and disturb the CO
2
regulator
system.
Please follow these safety precautions when changing the filter:
•
Always use the original filter (contact Esco Medical or your local distributor for more
details or ordering).
•
Change the filter once every year.
•
Failure to change the filter on time will result in low/no cleaning of incoming CO
2
gas.
•
Warranty void if wrong/ non-original filter is used.
Please refer to the service manual for replacement instructions.
37.14 Internal 0.2µ filter for N
2
gas
The round-shape in line 0.2µ HEPA filter for N
2
gas removes any particles found in the
incoming N
2
gas. Failure to use the HEPA filter may cause damage to the high precision flow
sensor, calculate the amount of N
2
gas entering the system, and disturb the N
2
regulator
system.
Please follow these safety precautions when changing the filter:
•
Always use the original filter (contact Esco Medical or your local distributor for more
details or ordering).
•
Change the filter once every year.
•
Failure to change the filter on time will result in low/no cleaning of incoming N
2
gas.
•
Warranty void if wrong/ non-original filter is used.
Please refer to the service manual for replacement instructions.