Error #4
This error indicates that at least one of the batteries has reached the maximum
temperature allowed for its working conditions.
MyLab
displays the following
message and shuts down automatically:
Should this situation occur, contact Esaote Service.
Power Supply Error Messages
Whenever an error in the management of the power supply occurs,
MyLab
displays a numbered error message: the number in the warning message
indicates the type of error.
Error #5
This error indicates an overheating problem of the power supply.
MyLab
displays the following message:
If this situation occurs, shut down
MyLab
and leave it off for a while. Verify
that there is adequate ventilation to prevent the overheating of the device.
Should the problem persist, contact the Esaote Service department.
Error #6
This error indicates that a fan is not working.
MyLab
displays the following
message:
If this situation occurs, press
OK
and then shut down
MyLab
. Verify that
nothing is blocking the fan functioning, especially on the rear panel.
Should the problem persist, contact the Esaote Service department.
Error #7
This error indicates that a fault of the internal voltages occurs.
MyLab
displays
the following message:
Error #4: problem with battery status.
The automatic shut down will start in a
few seconds.
Error #5: overheating!
Please, contact the Service department.
Error #6: problem with fan.
Please, contact the Service department.
Error #7: problem with internal voltage.
Please, contact the Service department.
Summary of Contents for MyLab SigmaVET
Page 1: ...Rev 10 June 2021 MyLabSigmaVET GETTING STARTED 350037700...
Page 2: ...MyLab G E T T I N G S T A R T E D ii...
Page 5: ...MyLab G E T T I N G S T A R T E D v GETTING STARTED RED Declaration of Conformity...
Page 6: ...MyLab G E T T I N G S T A R T E D vi...
Page 12: ...I N D E X MyLab G E T T I N G S T A R T E D xii...
Page 120: ...P E R F O R M I N G A N E X A M MyLab G E T T I N G S T A R T E D 7 10...