18 Repair
44
17
D
ATA EXCHANGE WITH
PC
To exchange data with PC, use the dock station, interface cable (please refer to
clause 12) and software.
Software and user guide are available at
www.eriskip.com or contact
the Technical Support Service of the manufacturing company
If the gas detector was used or placed for storage at a negative
temperature, hold the gas detector within 2–4 hours at a temperature
from 0 to +30 °С before connection to PC.
Functions that could be changed using software:
Switching on/off zero self-calibration of each sensor;
Change of ALARM 1 and ALARM 2 values;
Setting of date and time;
Setting of archiving events;
Adjustment of gas detector;
Verification of gas detector;
Calibration of gas detector;
Configuration of measuring channels from sensors library;
Setting of calibration concentration values;
Switching on/switching off additional functions;
Blocking preventing the gas detector switching off out of the dock station;
Blocking preventing the gas detector calibration using menu (calibration only
using software and PC)..