Safety notes
Precautions during operation/use
•
Before using the unit, make sure that the medical product consultant has conducted an introduction
with regard to function and safety precautions.
•
The unit must not be used if there are any doubts as to it being isolated from earth or the suitability of
the mains cable.
•
The unit must only be used with a mains cable that complies with the regulations for use in medical
technology.
•
The unit is not intended for use in wet rooms, outdoors or in areas where there is danger of explosion.
•
Before initial operation, the device must be adjusted using the base adjustors at the back to grant ab-
solute stability.
•
When the saddle is exchanged, make sure that the screws at the saddle are tightened hard enough so
that the saddle cannot be moved on the saddle pipe.
•
To move the handle bars and saddle, loosen the clamps and retighten them well afterwards. It is rec-
ommended to position the clamps with the levers positioned downwards. Secure clamping is ensured
when the clamps are again turned to this position after every adjustment.
•
When the patients mounts/dismounts, the therapist/physician needs to make sure that the patient's
feet do not get trapped in the pedal straps.
•
The holding straps on the pedals must fit perfectly across the upper side of the shoe and be fastened
with a Velcro strap.
•
For patients who weigh more than 140 kg or are taller than 190 cm, the separately available tilting
protection is recommended.
•
For patients weighing more than 160 kg, the additional tilting protection is mandatory.
Safety precautions when operating with other devices
•
When several devices are coupled, there is a risk that the leakage currents may add up.
•
The RS-232 interface, which can be used for communication with other devices, is galvanically isolated
to e
nsure the patient’s safety.
•
External devices must only be connected using an interface cable supplied by the manufacturer.
•
Portable communication devices, HF radios and devices labelled with the symbol
(non-ionic elec-
tromagnetic radiation) can affect the operation of this device (see section 0).
Precautions during maintenance
•
The device must be turned off and the power plug disconnected before cleaning with liquid cleaning
agents.
•
Only use standard cleaning agents for plastic surfaces.
•
The unit may only be opened, repaired and serviced by authorised and trained personnel. If the unit is
opened inadmissibly, the warranty becomes void.
Interference
The unit meets EMC regulations for medical products to ensure protection against emission and radiation.
Special caution needs to be taken when using this unit in combination with high-frequency devices (see sec-
tion 0).