Appendix
74/84
16.1.3
Table 204: Guidelines and manufacturer's
declaration – Electromagnetic interference
The Tanning system has been designed for use in the electromagnetic envi-
ronment specified below. The device operator must ensure that it is used in
such an environment.
Interference
immunity tests
IEC 60601 -
Test level
Defined
level
Electromagnetic environment – Guidelines
Conducted
HF noise
IEC 61000-4-6
Radiated
HF noise
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2,5 GHz
3 Vrms
3 V/m
Portable and mobile radios are not to be used at
shorter distances to the Model-No.: JK 98 / 48-0 and
its cables other than the recommended safe dis-
tance, which has been calculated according to the
transmitting frequency are made.
Recommended safe distance:
80 MHz to 800 MHz
800 MHz to 2,5 GHz
P is the transmitter's maximum output power (nomi-
nal rating) in Watts (W) according to the details pro-
vided by the manufacturer of the transmitter and the
recommend safe distance is given in metres (m).
The field strength of the stationary radio transmitter is
for all frequencies
b
in accordance with tests that were
carried out onsite
a
less than the determined level.
Interference is possible in the immediate surrounding
of equipment that carry the following symbols.
Comment 1: The higher value applies to 80 MHz und 800 MHz.
Comment 2: These guidelines might not apply to all situations. The propaga-
tion of electromagnetic waves is affected by absorption and reflexes caused
by buildings, objects and people.
a The field strength of a stationary transmitter e.g. standard cell phone stations
(mobile / cordless phones) and public mobile radio services, amateur stations,
MW and UHF radio and TV transmitters) cannot be precisely determined for the-
oretical reasons. It is recommended that an onsite test should be carried out in
order to determine the electromagnetic environment caused by stationary HF
transmitters. If the field strength of the massage device determined at the instal-
lation site exceeds the defined HF level listed above then the device must be
monitored with regard to its normal operation. If any abnormal characteristics are
recorded then it will be necessary to introduce additional measures, such as re-
orientation or converting the device.
b The field strength must be less than 3 V/m over the frequency ranges from
150 kHz to 80 MHz.