Cleaning and maintenance
10/2013
1013688-00
Operating instructions – 45/74
–
Open the clamps.
–
Lift the filter panel, pull it out of the guide and remove it.
–
Unplug the UV lamps.
–
Clean the reflector.
–
Only utilize original UV lamps.
–
Only touch new lamps in the handling zone.
–
Test the function of the safety switch.
The safety switch ensures that the device switches off in the event of a
faulty filter panel.
–
Press the switch.
If the switch springs back automatically then it is functional.
–
Replace the filter panel.