Cleaning and maintenance
48/68
4.8.1
Changing the chip card
DANGER!
Danger of being burnt by an over-powered lamp!
If the old chip card is left in the device, the new lamp set will
be run at too high power and the user may suffer heavy
burns!
–
The enclosed chip card must always be used when
changing the UV low pressure lamps!
The chip card belongs to the long UV low pressure lamps fitted in the canopy,
the side panels and the device's base.
The chip card is located in the 'Part support control'.
–
Opening the bed base: see page 46.
–
Remove the chip card belonging to the old lamp set.
–
Insert the new chip card with the arrow pointing downwards.
–
Close the bed base. You should hear the bed base click into place.