2  Cleaning and Disinfecting the Device Components

10

 

Cleaning and Disinfecting ergoline Medical Devices

2.7  ECG  Electrodes / Suction  Leads

2.7.1 Cleaning

ECG electrodes and suction leads should be cleaned after 
each use to ensure optimal hygiene and durability.

Take care not to scratch or damage the surface of the 
suction electrodes.

The number of cleaning cycles under normal operating 
conditions does not adversely affect the service life. 
Thorough cleaning ensures constant, high quality signal 
transmission and extends the lifetime.

Only a soft, dry or moistened cloth with one of the 
cleaning agents listed in section 1.1.1 should be used for 
cleaning.

Notice

•  Strictly adhere to the manufacturer's instructions 

for use of the substances, e.g., regarding dosage 
and exposure times.

Warning

 

  Equipment Damage

•   Use only cleaning agents and disinfectants 

recommended by ergoline.  Unsuitable products 
may damage the device.

•  Do not, under any circumstances, immerse the device in 

liquids or allow liquid to enter the interior of the device.

2.7.2 Disinfection

To ensure complete disinfection, the manufacturer's direc-
tions must be strictly observed, the prescribed exposure 
times being of particular importance.

The parts should be disinfected after each use with one of 
the disinfectants listed in section 1.2.1.

•  Spray the disinfectant onto the exterior of the suction 

leads and the suction cups. Observe the exposure time 
specified by the manufacturer.

•  In the case of heavy soiling of the suction leads and 

suction cups, disinfect their exterior with disinfection 
wipes or impregnated cleaning tissues.

•  Spray the disinfectant on the inside of the electrodes. 

Observe the exposure time specified by the manufacturer.

•  Disinfect the sealing lips and electrode surfaces with 

disinfection wipes or impregnated cleaning tissues.

Summary of Contents for ergoselect 10 M

Page 1: ...Cleaning and Disinfecting ergoline Medical Devices 201000641000 Version 2020 10 05 Rev 00 English...

Page 2: ......

Page 3: ...s for specific products are available for down load at the ergoline website www ergoline com All trademarks appearing in this document are trademarks of their respective owners Their protection is ack...

Page 4: ...2 Cleaning and Disinfecting ergoline Medical Devices...

Page 5: ...ays of Control Terminals 8 2 5 Upholstery 8 2 6 Blood Pressure Cuff 8 2 6 1 Removing the Microphone 8 2 6 2 Cleaning 8 2 6 3 Disinfection 9 2 6 4 Inserting the Microphone 9 2 6 5 Inspection and Functi...

Page 6: ...4 Cleaning and Disinfecting ergoline Medical Devices...

Page 7: ...products containing any of the following active ingredients is prohibited Ethanol 90 Acetone Hexane Scouring powder Substances harmful to plastic Organic solvents Ammonia based cleaning agents Abrasiv...

Page 8: ...1 Cleaning Disinfection General Information 6 Cleaning and Disinfecting ergoline Medical Devices 1 2 2 Unapproved Disinfectants 100 alcohol Virex HB Quat Conventional detergents Conductive solutions...

Page 9: ...necessary a domestic non caustic cleaning agent or a 70 alcohol solution can be used for grease and finger marks Immediately wipe off all excess cleaning agent Wipe the device surface down with a clo...

Page 10: ...e upholstery down with a soft cloth moistened with soap water The cloth should only be moist and not dripping wet If the cleaning agents and disinfectants used are caustic or contain alcohol they may...

Page 11: ...disinfectants listed in section 1 2 1 Notice Strictly observe the manufacturer s instructions for use 2 6 4 Inserting the Microphone Slip the microphone into the pocket the metal side facing the arm G...

Page 12: ...e Use only cleaning agents and disinfectants recommended by ergoline Unsuitable products may damage the device Do not under any circumstances immerse the device in liquids or allow liquid to enter the...

Page 13: ...dehydes alcohol amines alkalis or quaternary ammonium compounds are suitable for cleaning and disinfecting the SpO2 sensors if the combination and concentration of active ingredients is similar to the...

Page 14: ...in the connector and socket must be repeatedly flushed and drained 6 Make sure that the sensor has been throughly cleaned without any residue 7 Repeat the entire manual cleaning process if necessary...

Page 15: ...the Device Components Cleaning and Disinfecting ergoline Medical Devices 13 2 9 Ear Clip Wipe the ear clip with a cloth moistened with 70 isopropyl alcohol After cleaning wipe the ear clip with a clea...

Page 16: ...201000641000 Version 2020 10 05 Rev 00 English 14 Cleaning and Disinfecting ergoline Medical Devices...

Page 17: ......

Page 18: ...ergoline GmbH Lindenstra e 5 72475 Bitz Germany Tel 49 0 7431 98 94 0 Fax 49 0 7431 98 94 128 e mail info ergoline com http www ergoline com...

Reviews: