background image

15 

 

 

Lifting mechanism: 

Actuator runs both ways. Actuator is mounted correct. Lifts from 

bottom to top with max SWL 185 kg. (408 lb). Ratchet works when lowering. 

 

Hand Control: 

All functions work as specified in the user guide. 

 

Control Box and battery: 

Correctly mounted. No visible defects. 

 

Wheels: 

Wheels and brakes work. Securely fastened. 

 

Frame and Welding’s: 

Must be checked for visual signs of cracks, rust or 

deformations. 

 

Handles: 

Securely mounted without backlash. 

 

Service interval:

 After service has been carried out the service interval must be reset. 

Warranty - 

See sales and delivery terms for warranty information a

www.Winncare.com

 

5.2

 

Troubleshooting 

Fault:

 

The hoist does not 
react when the hand 
control buttons 
UP/DOWN are used.

 

Possible cause: 

Emergency stop is activated. 

Solution: 

Release the emergency stop (see section 2.3)  

Possible cause: 

Defect hand control. 

Solution: 

Try using another hand control. Alternative is 
to get a new hand control from the dealer. 

Possible cause: 

Flat battery. 

Solution: 

Charge battery (see section 2.4). 

 

Fault:

 

Some functions work, 
while others don’t. 

Possible cause: 

Fault in the control box. 

Solution: 

Replace the control box. 

Possible cause: 

Defect hand control. 

Solution: 

Try using another hand control. Alternative is 
to get a new hand control from the dealer. 

 

Possible cause: 

Fault in electrical circuit  

Solution: 

Check wiring  

Fault:

 

The lifting actuator is 
noisy or vibrates 
abnormally 

 

Possible cause: 

Lifting actuator is defect. 

Solution: 

Replace the lifting actuator. 

 

 

Fault:

 

Does not charge

 

Possible cause: 

Charger is not connected to the main supply. 

Solution: 

Connect charger. 

Possible cause: 

Defect control box. 

Solution: 

Try another control box. Order a new from 
Winncare Nordic or the dealer. 

Possible cause: 

Defect battery. 

Solution: 

Replace battery. 

 

Summary of Contents for Solar/185

Page 1: ...on Important information This user instruction must be read before the Solar 185 is used This user instruction is intended for technicians installers therapists and other personnel in connection with...

Page 2: ...2...

Page 3: ...3 Safety functions 11 2 4 Charging 12 2 5 Transport 13 3 Use 13 3 1 Using the Solar 13 3 2 Using the control box 13 4 Maintenance 14 4 1 Cleaning 14 4 2 Storage 14 4 3 Special environments 14 5 Check...

Page 4: ...cific information about which models can be used The Solar is intended for indoor use under normal temperatures 2 C to 40 C 36 F to 104 F and relative humidity 40 90 1 2 Safety instructions Important...

Page 5: ...d the hand control If the checks or the use of the Solar generate unusual noises then the Solar must not be used before a qualified technician has carried out a safety check on the Solar Check the lif...

Page 6: ...pair or annual check Content of the box 1 Base with legs and leg spreading 2 Mast with actuator and lifting arm 3 Spreader bar 4 Hand control 5 Control box 6 Handles 7 Battery 8 Tools Allen key spanne...

Page 7: ...the bolt nut Fig 1 6D Put the cap on the nut Fig 1 6D 4 Assembly the handles on the mast with two screws and tighten with the Allen key Fig 1 6E Fig 1 6E 5 The hand control lifting arm and electrical...

Page 8: ...ox Fig 1 6G 7 Attach the spreader bar Rotate the attachment point so it is pointing upwards insert the spreader bar and rotate 90 clockwise while pressing down firmly Fig 1 6H Note A firm press must b...

Page 9: ...der bar Winncare Nordic accepts no responsibility if slings are attached incorrectly it is always the onus of the carer helper to ensure that the sling is correct for the situation needs and is correc...

Page 10: ...Connecting plug electrical leg spreading 4 Connecting plug power supply 5 ON indicates that main supply is on 6 Charge indicates charging on the battery see section 2 4 7 Lift up down The up down func...

Page 11: ...reading Press and the leg spreading moves together Press and the leg spreading will spread outwards 2 3 Safety functions Emergency stop This function is only to be used in an emergency e g if the Sola...

Page 12: ...ere are 3 symbols on the hand control which indicate the power level in the battery Battery 25 An audible battery alarm will alert when there is a 25 capacity remaining in the battery Battery 50 Recom...

Page 13: ...e counters Battery state Overload information Service visit reports To access this information a readout tool is required This service data tool can be purchased from Winncare Nordic or your local dea...

Page 14: ...lease contact your local dealer or Winncare Nordic 5 CHECK SERVICE DATA AND TROUBLESHOOTING 5 1 Annual check The manufacturer Winncare Nordic stipulates that a qualified and authorised person MUST che...

Page 15: ...Release the emergency stop see section 2 3 Possible cause Defect hand control Solution Try using another hand control Alternative is to get a new hand control from the dealer Possible cause Flat batte...

Page 16: ...ty cycle 10 max 2 min 18 min Manually emergency lowering Yes Electronically lowering Yes Electrical details Power supply Input 100 240 VAC 50 60 Hz max 0 4 A Output voltage 24V Control box Output 250...

Page 17: ...gs spread 360 mm 14 in g Max reach at 600 mm reference height 580 mm 23 in h Max reach from base 600 mm 24 in i Reach from base with legs spread to 700 mm 180 mm 7 in j Base height 120 mm 5 in k Floor...

Page 18: ...nvironment if not handled in an environmentally responsible way 7 2 Battery The Solar is equipped with one battery which may be returned either to the local recycling station or to Winncare Nordic at...

Page 19: ...r bar 2 points 48 cm spreader bar 3 points spreader bar Spreader bars with weigh scale 017 00117 017 00144 017 00143 2 points 58 cm spreader bar with weigh scale 2 points 48 cm spreader bar with weigh...

Page 20: ...3 4 2 E F C L A S S I E N I S O 1 0 5 3 5 Phone 45 70 27 37 20 Fax 45 70 27 37 19 Email Info winncare dk www winncare com Manufacturer Winncare Nordic ApS Taarnborgvej 12 C 4220 Korsoer Denmark 910 00...

Reviews: