© Erbe Elektromedizin GmbH
Contacts worldwide: see erbe-med.com
1. Wait until the ERBEJET 2 display shows
Prime
.
2. Press the
START
selection button.
3. Verify whether the suction hose of the
pump cartridge fills with the separating
medium. The priming process takes up to
20 seconds.
4. Functional test: Activate the footswitch
until separating medium flows from the
tip of the instrument.
Prerequisite: The option is switched on (can be
carried out by Erbe employees).
1. Press the Selection button next to the
menu item
Program / Setup
.
2. Press the Selection button next to the
menu item
Info
.
Note:
If you activate the ERBEJET 2 frequently,
you can reduce possible obstructions in the
instrument (e.g. in the HybridKnife).
The pump cartridge is intended for treating
one patient. When treating this one patient,
you can change the instrument as often as you
like.
Press the release button to detach the in-
strument.
Note:
Use aseptic techniques when you
change the instrument.
Before you switch off the device:
1. Press EJECT to remove the pump car-
tridge. Observe the instructions on the
display.
2. Remove the pump cartridge.
Operating steps for ESM 2, only
for applicator and monopolar ap-
plicator:
Switch on the ESM 2.
Connect the vacuum hose with the mem-
brane filter.
Note:
Use a new membrane filter in the fol-
lowing cases: For each new application. If the
membrane filter has been unpacked for longer
than 7 days. If the membrane filter has
absorbed moisture.
1. Attach the adapter onto the connection
port for the suction hose.
2. Attach the applicator suction hose to the
adapter.
10
Priming the pump cartridge
11
Displaying used separating
medium
12
Reduce obstructions
13
Changing the instrument
14
Removing pump cartridge
15
Switching on ESM 2
16
Connect membrane filter with
vacuum hose
17
Connecting the vacuum hose
18
Inserting the suction bag
19
Attaching the suction hose to the
suction container
20
Changing the suction mode