80114-381_V23323
2020-07
HYDROSURGERY
QUICK GUIDE
ERBEJET
®
2
This quick guide is no substitute for user manuals and notes on use. This quick guide does not contain any safety instructions. Familiarize
yourself with the safety instructions in the user manuals and notes on use.
For all instruments:
Switch on the ERBEJET 2 unit.
1. Use the selection buttons
to select a
program for your application.
2. Confirm using the
OK
selection button.
Connect the fluid tube connector and
pump cartridge. The connector must audi-
bly click into place.
Note:
For sterile procedures, the connection
should take place in a sterile area.
1. Wait until the ERBEJET 2 display shows
Please insert new cartridge!
2. Slide the pump cartridge into the pump
shaft. The pump cartridge must audibly
click into place.
Only for HybridKnife, HybridAPC
probe and monopolar applica-
tor:
1. Switch on the VIO.
2. Select a program for your application.
Connect the return electrode to the VIO.
Insert the HF connector into a suitable
socket (see above).
Insert the FiAPC connector into the APC
socket.
For all instruments:
1. Connect an infusion set to a bottle or a
bag with separating medium. The sepa-
rating medium is sterile, physiological
saline solution.
2. Hang the bottle or the bag.
Note:
Hang the bottle or the bag a maximum
of 1 m above the standard rail. The lowest
possible attachment point is at the height of
the standard rail.
3. Open the roller clamp.
4. Fill the infusion set until it has been fully
primed.
5. Remove the protective cap from the pump
cartridge.
6. Connect the pump cartridge and infusion
tube.
!
1
Switching on ERBEJET 2
2
Selecting the program
3
Connecting the fluid tube
connector and pump cartridge
4
Inserting the pump cartridge
5
Select program on VIO
6
Connecting the neutral electrode
7
Connect the HybridKnife or
monopolar applicator to the VIO
8
Connect the HybridAPC probe to
the APC unit
MO /
MONOPOLAR
MF-U
APC
9
Connect the separating medium to
the pump cartridge