Initial Qualification Test ERAVAP
Acknowledging the following points and completing the
instrument performance check
ensures
that the
analyzer was installed successfully
and is fully operational.
By completing this form it is made sure that the
operator was instructed on all essential
parts of its operation
and that the analyzer meets our highest quality standards at the time
of installation.
Instrument Model:
____________________
Serial No.:
EVP _ _ _ _ _ _ _ _
Software version:
8 _ _ _
Information
Received manual and operations video
_____________
Received test report
_____________
Informed about cleaning when crude oil or other higher
viscous samples are tested (see section “Cleaning”)
_____________
Informed about handling of samples with high vapor pressure
(e.g. LPG, live crude oils)
_____________
Instrument performance check
Perform a reference measurement using ASTM D5191 / EN13016-1
or
ASTM D6378
or
ASTM
D6377 with n-
Pentane (purity ≥ 99.5%)
or
ASTM D6897 with n-
Butane (purity ≥ 99%) and fill
in the actual measured values:
ASTM D5191, EN13016-1 & ASTM D6377:
Pentane (cooled to 0°C and air saturated): P
tot
= 112.8 kPa ± 1.2 kPa
____________ kPa
ASTM D6378:
Pentane: VP4 (P
abs
) = 107.9 kPa ± 1.2 kPa
____________ kPa
ASTM D6897:
Butane: VP
tot
(37.8 °C, V/L 0.5:1) = 356.6 kPa ± 6.0 kPa
____________ kPa