Electrical
Note:
Check the label on the back of the scanner for voltage information.
Environmental
Temperature:
Operation
5° C to 35° C (41° F to 95° F)
10° C to 32° C (50° F to 89° F)
when using the ADF
Storage
–25° C to 60° C (–13° F to 140° F)
Humidity:
Operation
10% to 80%, without condensation
Storage
10% to 85%, without condensation
Operating conditions:
Ordinary office or home conditions.
Extreme dust should be avoided.
Operation under direct sunlight or
near a strong light source should be
avoided.
Note:
Specifications are subject to change without notice.
100
−
120 V model
220
−
240 V model
Input voltage range
AC 90 to 132 V
AC 198 to 264 V
Rated frequency
50 to 60 Hz
Rated current
0.5 A
0.3 A
Power consumption
Approx. 25W (operating)