Agilent 8700 LDIR Chemical Imaging System 

Long Term Shut Down and Start Up 

Instructions

 

 

 

 

Shut down the system 

To shut down the system:  

1

 

Save any changes to sample files. 

2

 

Click Eject and slide open the 8700 LDIR sample door to remove the sample.  

3

 

Close the instrument sample door. 

4

 

Close Agilent Clarity software. 

5

 

Turn off the instrument using the power button on the front of the instrument.  

The LED indicator on the power button will turn yellow. 

 

 

 

Figure 1 

Front of the 8700 LDIR

 

6

 

Switch off the rear power switch to disconnect the instrument from the mains power source. 

The LED indicator will be off. 