Important Information About Your New Meter
EasyMax V
blood glucose meter is designed and approved for testing fresh capillary blood from your
fingertip, palm or forearm. The meter is for outside the body (in vitro) use. It should not be used to
diagnose diabetes.
Only use
EasyMax V
blood glucose meter with
EasyMax
Blood Glucose Test Strips. Other test strips
will give inaccurate results.
Testing is not valid for neonatal blood specimen.
Do not disassemble the meter as this may cause damage to the components resulting in incorrect
reading. Disassembling the meter will also void the warranty.
Always keep the meter clean and store it in a safe place. Protect the meter from direct sunlight to ensure
a longer lifespan.
You should not store the meter or test strips in a car, a bathroom or a refrigerator.
Keep the meter, test strips and lancing device away from children and pets.
You should not test critically ill patients with home-use blood glucose meters.
Elevated levels of acetaminophen, uric acid, gentisica acid, levodopa, dopamine, ascorbic acid and
methyldopa may affect results.
Incorrect results may occur when performing the test. If you believe you are not feeling well, please
contact your healthcare professional.
Summary of Contents for EasyMax V
Page 48: ...48...