background image

Quick Guide

 

▼ Time Mark

 

▼ Enable Flight Mode

 

▼ Wake Up

 

▼ Repeat Start

6> AccessKey

You could enable and apply 

AccessKey (Device Password)

 to data logger after connecting to logger 

via App or during configuration, once applied, user will be required to provide correct AccessKey when 
connecting to or programming data logger.

  shows on LCD once AccessKey was enabled.

Protection level determines what actions user could take when connecting to an AccessKey protected 
data logger through Bluetooth communication.

1 (Normal)

,  AccessKey is not required when connecting to and downloading an AccessKey protected 

logger, but AccessKey is required when performing further control actions like configuring, starting or 
stopping logger.

2 (High)

,  AccessKey is required when connecting to an AccessKey protected logger.

Press 

 button for 3 seconds to add a time stamped mark

Up to 9 marks available.

If device was set to 

Repeat Start 

during programming, then user could press 

 to start a new 

recording again directly after recording stops, no need to program device again. 

You could let Tempod Bluetooth logger enters into flight mode in occasions where radio-frequency 
signal transmisstion is not allowed. Connect to Tempod bluetooth logger via TempSen Go App, and then 
enable 

Flight

 mode in 

Device Connected

 page. Logger in flight mode will suspend Bluetooth signal 

transmisstion, so it will not be found or connected by TempSen Go App, but recording continues. User 
could have logger quit Flight Mode by turnning on Bluetooth on logger manually or wake logger up 
remotely.

You could wake logger up from flight mode by using 

Wake Up 

function. Tap 

Wake Up

 at bottom bar 

of 

Devices 

page on 

TempSen Go 

App, TempSen Go will automatically wake up 

ALL

 nearby TempSen 

Bluetooth loggers in flight mode, loggers waken up will turn on Bluetooth service.

Press and hold 

 button for 3 seconds to manually stop recording,  meanwhile 

 (stop) 

flashing on LCD reminds you to keep current button action until 

 becomes static, recording stopped 

successfully. You could also stop recording through 

TempSen Go App

, tap 

 at bottom bar in 

Device 

Connected

 page to stop recording.

5> Download Data (3 options)

4> Stop Recording

Connect Tempod

®

 logger to USB port, logger begins to generate PDF report automatically, and it would 

take couple of seconds to complete. The PDF report will be available in logger created removable 
storage device "TempSenDisk". Please copy the PDF report and save it to your computer. 

Connect to Tempod

® 

bluetooth logger via TempSen Go App, and then tap 

 on bottom bar in 

Device 

Connected

 page to download data from logger. Data downloaded will be saved in local database on 

your smart phone, you could view all stored data files in 

Data Files

 pages, tap   on title bar of App 

screen to expand main menu, in which please tap 

Data Files

 to enter into 

Data Files

 page.

You could also download Tempod using TempCentre software, connect  device to computer, open 
TempCentre software, and then TempCentre starts searching USB data logger and then downloading 
data  automatically.

•  Please 

DO NOT

 perform any "Save" or "Save as" actions within any PDF reader 

software, otherwise content of PDF file will be reorganized by PDF reader and raw data 
embedded in PDF file becomes unrecognizable.

•  Please 

DO NOT

 remove Tempod from USB port while it's still in process of generating 

PDF report.

 

▼ Protection Levels

•  Please note logger with remaing battery of one bar (or lower) or in Stopped or Error 

status could not enter into Flight mode. 

•  Please note logger in flight mode consumes more power than normal mode and  

shorten battery life as result , so please 

DO NOT

 use flight mode frequently.

 

Generating PDF 

 

Connecting to USB 

 

▼ Download PDF Report

 

▼ Download Data via TempCentre

 

▼ Download Data via TempSen Go App

7> Other Actions

8> Flight Mode

•  AccessKey will be automatically synced to cloud after user logged into TempSen Cloud 

on TempSen Go App, and users in the same organization with permission of Share 
AccessKey will receive AccessKey automatically when connecting to this logger.

•  Please contact us by sending email to [email protected] if Accesskey was lost.

•  Please note Wake Up function NOT applicable on all mobile devices.

  shows on LCD after flight mode was enabled, meanwhile   icon disappears.

Press and hold  

  button for 3 seconds to activate device and start recording. 

Meanwhile  

  (go) flashing on LCD  reminds you to keep current button action.

3> Start Recording

 

Start Delay

 (12.5 days)

 appears on downleft corner of LCD indicating recording begins.

LCD changes to time countdown if device was set to a start delay after 
successful activation.

You could also start recording via 

TempSen Go App

 on your mobile device, tap 

 at bottom bar in 

Device Connected

 page to start recording.