046-07-GA_Oxi_B-a_eng 12
For these reasons, avoid storing the OxiQuant B unnecessarily at high
ambient temperatures, in very dry environments or under increased partial
oxygen pressures.
6.3.
Changing the sensor
??
Undo the connecting nut on the sensor and remove the jack plug
??
Dispose of the sensor
??
Connect new sensor and check function
6.4.
Packaging and storage
During storage, the sensor consumes the oxygen from the space inside
the gas inlet opening. Ageing is therefore reduced if the sensor is stored
in the original packaging. However, before it is ready to operate, the
sensor requires some time after it has been removed from the packaging,
depending on the storage period and temperature, before its
measurement value becomes stable. This time can be up to 30 minutes.
Meter calibration should therefore be carried out after the sensor has
stabilised or should be repeated at that stage. A storage temperature
between 5 and 15°C is recommended to reduce the necessary
stabilisation time.
Storage:
-
Temperature range -20°C to 50°C / store in original packaging.
Labelling:
Item description: oxygen sensor
Application: determination of oxygen concentrations
Type: OOIXXX, XXX – consecutive type number
Manufacturer/supplier details:
Manufacturer:
E
NVITE
C
-Wismar GmbH, Alter Holzhafen 18, D-23966
Wismar, Germany
Telephone / Fax: +49 3841 360 1 / +49 3841 360 222
Sensor contains electrolyte and lead!