Environmental Instruments Fluoroprobe FL-2-S0 O & M Manual Download Page 5

 
 

Note: If the probe is being air calibrated in a pipe where the end of the pipe is 
submerged, 

vent holes must be drilled on both ends of the pipe to allow ambient 

air to purge the gases, which might give a false air calibration value.  Drill holes in 
pipe just above the water line and at the top of the pipe.  When air calibrating, the 
probe should be suspended between the holes in a white PVC pipe to help reflect 
heat from the sun. 

 
Do not air calibrate the FL-2 if air temperatures is + 10 or -10 degrees from water 
temperature.  Leave the probe in the water and use a handheld to compare readings.  The 
FL-2 has temperature compensation, so never calibrate it until you know the probe has 
had ample time to adjust to temperature.  If the readings do not look right to you 
investigate everything involved and wait till the next day to try again. 
 

Suggested calibration for wastewater treatment:

 There are four (4) basic ways to 

calibrate the Fluoroprobe

TM

 controller, which are described in the following sections.  

 
If you are using the FL-2 in a wastewater application and your operation range is 3 or 4 
ppm down to 0 ppm use a hand held instrument to compare or calibrate.  The sensor is 
more responsive and stable when left continuously in the water and calibrated within its 
normal operating environment.  Always clean sensor with a toothbrush and place back in 
aeration 30 minutes or longer before any calibration.  Bio-growth or debris covering the 
sensor will cause oxygen reading to slowly fall to zero, the only maintenance needed is 
once a month toothbrushing.  Many aeration basins never have this growth problem. 
 

1. Air Saturated Water Calibration:

. For this technique, a volume of water must be 

aerated for twenty (20) minutes or longer at a constant temperature.  The operator can 
use an aquarium pump and a stone diffuser (air stone).  A cylinder of some type is best. 
PVC pipe (2"or 3") with a cap on the bottom is a good vessel to use for this method. 
Place the probe in the aerated water keeping the probe sensor near the top of the 
container for maximum efficiency.  To obtain the correct current atmospheric pressure 
reading, use the local altitude or the true atmospheric pressure from a mercury 
barometer or weather bureau report.  The same air calibration table used for 
polargraphic or galvanic calibration may also be used for reference.  This technology is 
for long term monitoring and taking your time is very important when calibrating this 
instrument.  After the probe has been submerged for twenty (20) minutes or more in the 
air-saturated water, press the 

READ

 button on the faceplate of the instrument.  The 

operator will have about two (2) minutes to adjust the air calibrate adjustment screw on 
the faceplate to the air saturated water concentration.  No water flow or movement in 
the probe is necessary in order to obtain 0

2

 concentration readings.  A hand held 

oxygen meter can be used in calibration to assure air saturation.  If you are also zero 
calibrating, be sure to clean the probe before changing from one liquid to the next.  

Summary of Contents for Fluoroprobe FL-2-S0

Page 1: ...O M Manual MODEL FL 2 S0 NEW WEBSITE WWW thefluoroprobe com ENVIRONMENTAL INSTRUMENTS LLC 144 Industrial Park Drive SuiteA 3 Waynesville NC 28786 From Process Control Services Phone 02 9319 1808 Fax...

Page 2: ...ibration 3 Air Calibration 4 Calibration Using Hand Held DO Meter 5 Probe and Sensor Maintenance 6 Troubleshooting and Tips Figure 1 Installation Illustration Figure 2 Face Plate Descriptions Drawing...

Page 3: ...tional solar powered brush for cleaning the sensor can be ordered mounted on a float for remote or aquaculture applications Unlike polargraphic and Galvanic probes the FluoroprobeTM sensor does not co...

Page 4: ...ibrating The FluoroprobeTM is very stable and the longer it stays in the water the better The FluoroprobeTM is calibrated before shipping to a known zero 0 ppm standard and air saturated water 8 47 D...

Page 5: ...owth or debris covering the sensor will cause oxygen reading to slowly fall to zero the only maintenance needed is once a month toothbrushing Many aeration basins never have this growth problem 1 Air...

Page 6: ...t twenty 20 minutes Do not air calibrate if air and water temperatures are or 10o C difference Using the enclosed pressure altitude chart fig 3 and the altitude of the location where the unit will be...

Page 7: ......

Page 8: ......

Page 9: ...1 2 3 4 1 2 3 4 A B C D A B C D Title Size Number Rev Date File Name Drawn By Sheet of RELAY LOCATION RELAY 3 10A 250VAC RELAY 2 10A 250VAC RELAY 15A 250VAC TERMINAL POS WARRANTY ZERO LEAVE ALONE...

Page 10: ...ile Name Drawn By Sheet of THIS POT ADJUSTS THE 4mA AT 0 PPM THIS POT ADJUSTS THE SPAN CAN BE ADJUSTED FOR 20mA AT 10 PPM TO 19 99 PPM BLUE WIRE 4 20 WHITE WIRE 4 20 NOTE MAX LOOP RESISTANCE IS 550 oh...

Page 11: ...MODEL FL 2 SWING OUT WIRING DIAGRAM LINE NEU 4 20 4 20 Earth RELAY 1 RELAY 1 RELAY 2 RELAY 2 230 VAC RELAY 3 RELAY 3...

Page 12: ...0 2ppm Sensor Lifetime Based solely upon the application environment and the effect on Silicone Rubber 3 year Sensor warranty on approved applications 4 20 mA Max Resistance 550 ohms Features Tempera...

Page 13: ......

Reviews: