Environmental Express IAQ 15 Connect Operating Instructions Manual Download Page 11

11 | 

environmentalexpress.com

 

 

Section 10 -

 

Battery Status and Charging 

 

The battery level icon appears in the upper right-hand corner of the screen. A fully charged battery will have 5 

bars showing. As the battery level drops during use, the number of bars will decrease from right to left. NiMH 

batteries constantly work at regenerating themselves and releasing more power.  It is possible for the displayed 

battery level to fluctuate (increase or decrease) during use.

 

When the battery level becomes very low (1 bar), the red LED will start flashing. Some additional samples may 

still be taken, but the unit should be plugged into power source using the AC adapter as soon as possible. The 

pump may be operated using standard AC power supplied by the charger.  

If the battery’s voltage drops low enough that the unit will not operate reliably or if for any reason an 

inconsistent motor speed is detected, the unit will shut down. During shutdown, the amount of actual run time 

will be saved. When the power is restored, 

Incomplete Run

 will be displayed on the home screen. Previous 

sample may be continued, or a new sample may be started.  

To charge the battery pack, plug in the AC charger on the right side of the pump. The yellow LED will flash when 

the unit is charging. When charging is complete, the yellow LED will remain constant. A full battery charge takes 

approximately 3 hours. If the pump will not be used for an extended period, it will slowly lose its battery charge 

and will need to be recharged before use. During these extended periods, the pump can be kept on AC power so 

the battery receives a trickle charge. The pump can perform all sampling modes while charging. If charging and 

sampling or calibrating at the same time, the LED will alternate between flashing yellow and green. 

 

Section 11 -

 

Battery Replacement 

 

 

 

1.

 

Insert the battery cover key into the top of the battery cover on either side of the tripod mounting 

thread.  

2.

 

Pull out at the top of the battery cover and then remove the cover. 

3.

 

Remove battery pack from the enclosure to the side of the pump and then unplug cables. 

4.

 

Using replacement battery pack, plug in, and then place in enclosure and close battery cover.  

Take care not to pinch the battery cables with bottom of battery cover.        

 

        

            

 

 

1                             2                                         3 

 

 

Summary of Contents for IAQ 15 Connect

Page 1: ...T 1 800 343 5319 1 843 881 6560 W environmentalexpress com LA03115 Rev 1 IAQ 15 Connect Operating Instruction Manual...

Page 2: ...ncluded 5 Section 5 Physical Description 6 Section 6 Prior to Operation 7 Section 7 Calibration 7 Section 8 Sampling Modes 8 Section 9 Retrieval of Event Records with App 10 Section 10 Battery Status...

Page 3: ...ble for any consequential special indirect or compensatory damages arising from use of or in conjunction with this product The maximum liability of Environmental Express whether by reason of breach of...

Page 4: ...S s Operation is subject to the following two conditions 1 This device may not cause interference 2 This device must accept any interference including interference that may cause undesired operation o...

Page 5: ...in a manner for which it is not intended Technical Specifications Flow rate 15 LPM Accuracy 5 Housing PC ABS plastic Battery rechargeable nickel metal hydride NiMH 9 6 V 2500 mAh Run time 99 five min...

Page 6: ...ner 2 LCD 3 F1 button increase 4 F2 button enter 5 F3 button decrease 6 Bluetooth button 7 Power On Off and Wake up button 8 AC charger input jack Bottom View of Pump 1 Battery pack Battery cover 2 Gr...

Page 7: ...very day of use A flow indicator for Air O Cell cassettes is supplied with the pump for ease of calibration You may also calibrate the pump using an optional TSI 4046 Primary Calibrator with attachmen...

Page 8: ...tures a choice of 3 different sampling modes We strongly recommend that the sampling mode is chosen after calibration is complete Quick Sample This mode allows easy sampling with the most commonly use...

Page 9: ...llowed by one cycle off time Enter the Cycle Run Time run time will be the same for all run cycles Enter the Cycle Off Time off time will be the same for all off cycles 5 Press F1 button Run Time of C...

Page 10: ...10 environmentalexpress com 4 Once complete check mark will appear between the IAQ 15 Connect and mobile device with message stating Bluetooth Data Transfer Success...

Page 11: ...be displayed on the home screen Previous sample may be continued or a new sample may be started To charge the battery pack plug in the AC charger on the right side of the pump The yellow LED will fla...

Page 12: ...Section 13 Maintenance Proper care and maintenance of your IAQ 15 Connect pump is essential for a long operational life It is a delicate electronic device and should be treated accordingly Rough and a...

Page 13: ...art Heavy duty tripod sampling stand ZA0043 Remote extension tubing 5 ft ZBP 310 Calibration kit with TSI 4046 Primary Calibrator ZBP 200 CAL Flow indicator for calibrating Via Cell cassettes ZBP 305...

Reviews: