Envent Engineering Ltd.
Page 35
Revision 1
MAINTENANCE
The 330S-Ex and 330SDS-Ex H
2
S analyzer will provide reliable service with
very little attention. If the analyzer is kept clean there should be no
requirement to recalibrate from factory gain settings. However, regular
check-up (at least every three months) will ensure that the analyzer is
operating to specifications.
•
Ensure that the H
2
S sensing tape take-up and feed reels are tight
•
Ensure that the flow meters, humidifier tubing and sample chamber
tubing are free of liquid or particulate contamination. If the sample
conditioning system is flooded with liquid, refer to "Sample Conditioning
System Cleaning Procedure" on page 36.
•
Ensure there is enough H
2
S sensing tape, especially if a low H
2
S
sensing tape sensor is not installed. Refer to "H
Procedure" on page 35.
•
Check the sample conditioning filter(s) every time the H
2
S sensing tape
is replaced. Replace the filter(s) as required.
H
2
S Sensing Tape Change Procedure
Figure 17. 330S-Ex & 33SDS-Ex H
2
S Sensing Tape Change Procedure
Summary of Contents for 330S-Ex
Page 2: ......
Page 55: ...Envent Engineering Ltd Page 50 Revision 1...
Page 56: ...Envent Engineering Ltd Page 51 Revision 1...
Page 57: ...Envent Engineering Ltd Page 52 Revision 1...
Page 58: ...Envent Engineering Ltd Page 53 Revision 1...
Page 59: ...Envent Engineering Ltd Page 54 Revision 1...
Page 60: ...Envent Engineering Ltd Page 55 Revision 1...
Page 61: ...Envent Engineering Ltd Page 56 Revision 1...
Page 62: ...Envent Engineering Ltd Page 57 Revision 1...
Page 63: ...Envent Engineering Ltd Page 58 Revision 1...
Page 64: ...Envent Engineering Ltd Page 59 Revision 1...
Page 65: ...Envent Engineering Ltd Page 60 Revision 1...
Page 66: ...Envent Engineering Ltd Page 61 Revision 1 Safety Data Sheet for H2S Sensing Tape...
Page 67: ...Envent Engineering Ltd Page 62 Revision 1...
Page 68: ...Envent Engineering Ltd Page 63 Revision 1...
Page 69: ...Envent Engineering Ltd Page 64 Revision 1...