STACK 602
9 Maintenance
ENVEA
UK Ltd
(PC-000848-MA Issue 5, December 2021)
vii
Figure 6: Purge blower wiring.
4.4.3
Adjusting the Speed
The small purge blower unit has three speed settings: HIGH, MEDIUM, and LOW.
The typical amount of purge air required for
DynamicOpacity
®
sensors is between
30 and 60
l/min under normal stack conditions (typically ±10 mbarg) to both the
Transmitter and Receiver.
This can be achieved by setting the switch selection to
HIGH
). The
required speed is pre-set
ex-works
.
Figure 7: Speed selection for
DynamicOpacity
®
sensors.