11
C – Sensor Tab
If errors are found in the sensor by internal system tests, they are marked and the sensor and the
DRC display sensor error.
4.5.1.1 Basic parameters
A default parameter set is established for a new destination system with an empty EEPROM:
Parameter
Default
Meaning
ModBus address:
2
Sensor
1
DRC
TRH value:
1000
Current alarm threshold
TRH factor:
10
Factor
AutoSetup time:
5 [min]
time for AutoSetup function
Alarm delay:
2.5 [s]
No alarm until x seconds after the threshold isexceeded
Alarm hold:
10 [s]
The alarm is held for at least x seconds after the threshold value is exceeded.
Alarm hysteresis:
95 [%]
[%] Alarm cannot be regarded as able to be cancelled until it falls below
a figure of x percent of the threshold value.
Switch
AutoSetup time is fix
1
Fixed, not automatically extended AutoSETUP time = AS Time
0
With each new maximum value the Auto SETUP is extended
by the set AS Time
HW Switch S1 is enabled
1
S1 is enabled
0
S1 is ignored
DIN Rail relais is NC
1
DIN Rail relay is actuated as NC
0
DIN Rail relay is actuated as NO