To prevent an electrical shock hazard between the appliance and
other appliances or metal parts in close vicinity, an equalization-
bonding stud is provided. An equalization bonding lead must be
connected to this stud and the other appliances / metal parts to
provide sufficient protection against potential difference. The
terminal is marked with the following symbol.
ATTENTION
Consult accompanying
documents
Hazardous Voltage Present
Safety Class I
Equipment
Protective
Earth
Ground Symbol
Medical Equipment classified by Underwriters
Laboratories with Respect to Electrical Shock,
Fire and Mechanical Hazards only, in Accordance
with UL 60601-1 and CAN/CSA C22.2 No. 601.1.
Grounding reliability can only be achieved when equipment is
connected to an equivalent receptacle marked “Hospital Grade.”
ivNow-1
POWER REQUIREMENTS
230 V.A.C. — 50/60 Hz, 1 ph
0.15 kW, 0.7 Amps
Type B Equipment
BS 1363 Plug*
(UK only)
CEE 7/7*
220-230V Plug
ivNow-2
POWER REQUIREMENTS
230 V.A.C. — 50/60 Hz, 1 ph
0.3 kW, 1.3 Amps
Type B Equipment
BS 1363 Plug*
(UK only)
CEE 7/7*
220-230V Plug
ivNow-3
POWER REQUIREMENTS
230 V.A.C. — 50/60 Hz, 1 ph
0.45 kW, 2.0 Amps
Type B Equipment
BS 1363 Plug*
(UK only)
CEE 7/7*
220-230V Plug
ivNow-4
POWER REQUIREMENTS
230 V.A.C. — 50/60 Hz, 1 ph
0.6 kW, 2.6 Amps
Type B Equipment
BS 1363 Plug*
(UK only)
CEE 7/7*
220-230V Plug
ivNow-5
POWER REQUIREMENTS
230 V.A.C. — 50/60 Hz, 1 ph
0.75 kW, 3.2 Amps
Type B Equipment
BS 1363 Plug*
(UK only)
CEE 7/7*
220-230V Plug
ivNow-6
POWER REQUIREMENTS
230 V.A.C. — 50/60 Hz, 1 ph
0.9 kW, 3.9 Amps
Type B Equipment
BS 1363 Plug*
(UK only)
CEE 7/7*
220-230V Plug
*Other international plugs are available, contact factory for more information.
E L E C T R I C A L I N F O R M A T I O N
The power specifications are located on the unit identification nameplate. This nameplate
is permanently attached to the unit and must be located to verify power requirements.
P R E P A R A T I O N
Before operating the module(s), clean the exterior of the unit with a damp cloth and general hospital
cleaner (isopropyl alcohol)
D A N G E R
Ensure power source matches
voltage identified on appliance rating
tag. The rating tag provides essential
technical information required for any
appliance installation, maintenance
or repairs. Do not remove, damage or
modify the rating tag.
D A N G E R
Do not
use this warming appliance
in the presence of flammable
anesthetic mixture (with air or with
oxygen or nitrous oxide).
This could cause an explosion!
(Not category AP or APG equipment )
C A U T I O N
This unit has not been approved for
warming of blood or blood products.
WARNING
Injection Fluid manufacturer suggests not to
warm injection fluids ABOVE 40°C (104°F).
If fluids are warmed ABOVE suggested
temperature, they should be discarded.
Summary of Contents for ivNow Series
Page 19: ...MN 28929 REV 4 06 15 IVNOW MANUAL 17 ivNow 1 120V Wiring Diagram ...
Page 20: ...MN 28929 REV 4 06 15 IVNOW MANUAL 18 ivNow 2 120V Wiring Diagram ...
Page 21: ...MN 28929 REV 4 06 15 IVNOW MANUAL 19 ivNow 3 120V Wiring Diagram ...
Page 22: ...MN 28929 REV 4 06 15 IVNOW MANUAL 20 ivNow 4 120V Wiring Diagram ...
Page 23: ...MN 28929 REV 4 06 15 IVNOW MANUAL 21 ivNow 5 120V Wiring Diagram ...
Page 24: ...MN 28929 REV 4 06 15 IVNOW MANUAL 22 ivNow 6 120V Wiring Diagram ...
Page 25: ...MN 28929 REV 4 06 15 IVNOW MANUAL 23 ivNow 1 230V Wiring Diagram ...
Page 26: ...MN 28929 REV 4 06 15 IVNOW MANUAL 24 ivNow 2 230V Wiring Diagram ...
Page 27: ...MN 28929 REV 4 06 15 IVNOW MANUAL 25 ivNow 3 230V Wiring Diagram ...
Page 28: ...MN 28929 REV 4 06 15 IVNOW MANUAL 26 ivNow 4 230V Wiring Diagram ...
Page 29: ...MN 28929 REV 4 06 15 IVNOW MANUAL 27 ivNow 5 230V Wiring Diagram ...
Page 30: ...MN 28929 REV 4 06 15 IVNOW MANUAL 28 ivNow 6 230V Wiring Diagram ...