Combination Blanket/Fluid Warmer
Installation
WARNING
To prevent serious personal injury, death, or
property damage:
Do not steam clean, hose down or flood the
interior or exterior with water or liquid solution
of any kind.
Do not use water jet to clean. Failure
to observe this precaution will void the warranty.
DANGER
To prevent serious personal injury, death,
or property damage:
Do not use this warming appliance in
the presence of flammable anesthetic
mixtures (with air or with oxygen or
nitrous oxide).
Not category AP or APG equipment
Specifications:
• Dual-chamber warming appliance
• Stainless steel exterior and interior
• Doors are fully gasketed with a
magnetic closure
• WarmSafe™ heating system
incorporates a multiple zone
warming technology (Patent No: US
8,217,316; US 8,581,152) that heats
where and when it is needed . All
chamber surface temperatures are
monitored, providing an effi cient
balance of heat, low energy
consumption and minimal heat loss .
• Casters are standard
• Fluid chamber capacity: 30 1-liter
bottles or 24 1-liter bags
• Blanket chamber capacity: 30
blankets (16 .4 ft
3
)
Control:
• Adjustable temperature range
» Blanket chamber has an
adjustable temperature range
of 32°C - 71°C (90°F - 160°F)
» Fluid chamber can be
programmed to warm either
irrigation fl uids (IRR) or
injection fl uids (INJ), with
separate temperature ranges
provided depending on the
choice selected:
-
IRR temperature range:
32°C to 66°C (90°F to 150°F),
-
INJ temperature range:
32°C to 40°C (90°F to 104°F) .
• Accurate temperature
- IRR +0/-1 .67°C (-3°F)
- INJ +0/-1 .10°C (-2°F)
• Operate in Celsius or Fahrenheit
• LED display
• Built-in speaker for audible feedback
• Integrated controller lock-out
Additional features:
• Safety: The controller stores all
settings and resumes operation
using these settings in the event of a
power failure . When the power
is restore, the controller sounds an
alarm . The on/off status indicator
fl ashes .
• Safety: A warming shut-off system,
separate from the controller,
prevents overheating .
• Convenience: Access point at rear
panel allows the addition of
data logging or temperature
management hardware .
Clearance requirements:
4" (102mm) from rear
1" (25mm) from top
2" (51mm) from sides
1/2" (13mm) from bottom
Specifications and Features