Specification Information
Transportation and
storage conditions
▪
Ambient temperature range of -40°F to +159°F (-40°C to +70°C).
▪
Relative humidity range of 10% to 95%, non-condensing.
▪
Atmospheric pressure range of 7.25 psi to 15.37 psi (50kPa to 106kPa).
Operating conditions
▪
The warmer must acclimate to the room temperature it will be placed in—24
hours is recommended.
▪
The recommended environmental temperature range is 60°F to 90°F
(15°C to 32°C).
▪
The recommended relative humidity is above 20%, non-condensing.
Standards for
electrical equipment
▪
Medical equipment listed by Underwriters Laboratories with respect to
electrical shock, fire, and mechanical hazards only, in accordance with UL
61010-1 and CAN/CSA C22.2 No. 61010-1.
▪
Grounding reliability can only be achieved when the appliance is
connected to an equivalent receptacle marked “Hospital Grade.”
▪
IP-XO ordinary
Model
EC/DC 250
EC/DC 350
EC/DC 750
Intended use
Warming blankets
Weight
Net: 67 lb (30 kg)
Ship: 110 lb (50 kg)
Net: 91 lb (41 kg)
Ship: 140 lb (64 kg)
Net: 134 lb (61 kg)
Ship: 184 lb (83 kg)
Storage cavity
capacity
4–6 blankets
(2.5 ft
3
)
10–12 blankets
(3.5 ft
3
)
20–25 blankets
(7.5 ft
3
)
Temperature
range
90°F to 160°F (32°C to 71°C)
Clearance
requirements
3” (76mm) from rear
1” (25mm) from top
1” (25mm) from sides
3/4” (19mm) from bottom
E471516