a.

 

After setting up evaluation board for test connect RX input to PPG thru SMA 

AC coupling caps and RX output to load. 

b.

 

Load RX driving script and observe output on scope.   

c.

 

RX waveform control 

i.

 

Adjust Preemphasis magnitude.  Please check Preemphasis enable 

control box to activate. 

ii.

 

Adjust Preemphasis phase.  Please check Preemphasis enable 

control box to activate. 

iii.

 

Activate LNB by unchecking LNB buffer power down control.  Do not 

activate LNB while LIA is active. 

iv.

 

Activate LIA by unchecking LIA buffer power down control.  Do not 

activate LIA while LNB is active 

v.

 

 

 

3.

 

Gate Voltage DAC test 

a.

 

After setting up evaluation board for test connect VG pin to load 

Summary of Contents for ESI-1002

Page 1: ...Board User Manual Version 1 0 1 Introduction This document describes the basic connection and use for ESI 1002 evaluation board 2 Prerequisites 1 One 3 3V power supply 2 Input and Output SMA AC coupl...

Page 2: ...LAKESIDE DRIVE SUITE 228 SANTA CLARA CA 95054 www enspheresolutions com ESI 1002 Evaluation Board Schematic 1 Connect 3 3V power supply and ground to ESI board Power Supply Connections 2 Connect Chee...

Page 3: ...ONS INC 2900 LAKESIDE DRIVE SUITE 228 SANTA CLARA CA 95054 www enspheresolutions com 3 Turn on 3 3V power supply 4 Start GUI and Connect Please see GUI user manual for GUI operation Also see datasheet...

Page 4: ...ut to load b Load TX driving script and observe output on scope c IMOD amplitude control d TX waveform control i Adjust crossing percentage of waveform by changing Pulse width control ii Adjust Pre dr...

Page 5: ...g script and observe output on scope c RX waveform control i Adjust Preemphasis magnitude Please check Preemphasis enable control box to activate ii Adjust Preemphasis phase Please check Preemphasis e...

Page 6: ...ading in GUI 5 Run any generic test by setting registers manually If there is a particular test or setting you would like to run please contact Ensphere s applications team as they can advise as to ho...

Page 7: ...FOR A PARTICULAR PURPOSE Ensphere Solutions Incorporated assumes no responsibility for the consequences of use of such information or for any infringement of patents or other rights of third parties t...

Reviews: