EN109-1570750-41 IFU

 

6

 

Precautions 

 

 

It is important to read the following instructions carefully before using your ManuMax 
properly and safely. The manufacturer cannot be held responsible for the results of using 
this device for any purposes other than described in these operating instructions. 

 

 

  If the use of this device may have caused or contributed to an undesirable event such as death 

or serious injury to the user, the manufacturer AND the Competent Authority of the Member 
State MUST be notified immediately! 

  The use of accessories and cables other than those specified or supplied by the manufacturer 

of this equipment is not permitted. 

  The couch weighs approximately 100 kg and must be unpacked by at least two people. 
  Do not lift the couch by the table top sections or the all-around control. 
  Immediately after unpacking the couch, check whether the package contains all the items 

listed above, and check the couch for any damage. 
In case of damage, contact your local representative. Do NOT use the equipment! 

  Check the technical data section to ensure that the mains voltage meets the technical 

specifications of the couch. 

  Check the technical data to ensure that the environment meets the environmental conditions 

before using the couch. 

  Avoid exposure to direct sunlight, UV radiation, excessive dust, moisture, vibrations and 

impacts. 

  ManuMax treatment couches are not suitable for use in so-

called ‘wet rooms’ (hydrotherapy 

rooms). 

  Never move the couch with a person or a weight on the couch. Only move the couch on a level 

surface. The couch is not suitable for riding over thresholds. 

  The ManuMax treatment couches are intended to be used, and shall only be used, by or under 

the supervision of professional users in physiotherapy, rehabilitation and adjacent areas. 

  Never adjust a table top section when the patient is supporting on it with his/her full weight. 

The table top sections of some ManuMax couches are supported by gas springs. These ensure 
an easy adjustment and stable positioning of the table top sections. However, the gas springs 
are not strong enough to adjust a section while supporting the full weight of the patient. 

  The arm rests will not be completely blocked after releasing the gas spring. The locking is 

sufficient for supporting the arms, but is not suitable for exerting much more force on them. 
Please, do not sit on them. 

  It is very important for the life of the upholstery that it is properly cleaned. 
  Skin oils, perspiration, make-up products, hairspray, etc. and overly aggressive detergents and 

disinfectants can have a very detrimental effect on the lifespan of the upholstery. 

  Do not use any alcohol-based or aggressive cleaning agents. 
  If moisture or any material gets into the couch, turn it off, unplug it from the wall socket, and 

have it checked by an authorised service technician before using the couch again. 

  Never use a mix of disinfectants or a mix of disinfectants and soap. 
  Prevent moisture from accumulating in the hinges of the metal frame. (Prolonged exposure to 

moisture in the hinges may cause rust in the worst case scenario). 

  Regularly (monthly) check the functioning of the head/back section lock. 
  If the treatment couch is equipped with a motor, disconnect the plug from the wall socket if 

the height adjustment is not used for a prolonged period of time. 

  Have your couch revised annually by authorised Enraf-Nonius staff. 

 

 

 

 

Summary of Contents for MANUMAX

Page 1: ...Page 1 of 20 EN109 1570750 41 IFU MANUMAX EN109 1570750 41 IFU July 17 2019 Instructions For Use...

Page 2: ...IFU Table of Contents 1 Introduction 3 2 Symbols 4 3 Package Contents 5 4 Installation 6 5 Intended use and intended user 8 6 Precautions 9 7 Operation 12 8 Maintenance 15 9 Specifications 17 10 Conta...

Page 3: ...y of life of many people This Instructions For Use relates to the ManuMax Examination 2 2 the ManuMax Multi 3 3 and the ManuMax Multi 3 5 It is important to read the following instructions carefully b...

Page 4: ...e safely exposed Waste electrical items that can be recycled Indicates the electrical and electronic components of the device can be recycled and has to be disposed separately Manufacturer name addres...

Page 5: ...following is included with your treatment couch 1x 1570750 ManuMax User Manual PDF on CD 1x 1570751 ManuMax Information Booklet The use of accessories and cables other than those specified or supplie...

Page 6: ...on The couch weighs approximately 100 kg and must be unpacked by at least two people Do not lift the couch by the table top parts or the all around control Immediately after unpacking the couch check...

Page 7: ...he frame Hooks are fitted on the inside of the subframe to guide the power cord to one of the four legs Guide the cord to the most suitable leg depending on where the wall socket in your treatment roo...

Page 8: ...ded to support a patient in the desired position for examination or treatment within the physiotherapy rehabilitation and adjacent disciplines Intended User The ManuMax treatment couches are intended...

Page 9: ...ver thresholds The ManuMax treatment couches are intended to be used and shall only be used by or under the supervision of professional users in physiotherapy rehabilitation and adjacent areas Never a...

Page 10: ...couch is 250 kg This means that the total weight on the couch patient weight plus any other load such as manipulation by the therapist must not exceed 250 kg Never leave children alone with this devi...

Page 11: ...re not to make contact with the motor when placing the lift or emitter Placing objects under the couch with a height exceeding 95 mm may result in dangerous situations and serious damage Try to limit...

Page 12: ...o select the desired mode One or more beeps will sound The number of beeps represents the selected setting Press once 1 beep sounds lock enabled time of 30 minutes Press twice 2 beep sounds lock enabl...

Page 13: ...ax couches are supported by gas springs These ensure an easy adjustment and stable positioning of the table top sections However the gas springs are not strong enough to adjust a section while support...

Page 14: ...nlock the leg section lever 3 and put the leg section in the angled position using the other hand Never adjust a table top section when the patient is supporting on it with his her full weight The tab...

Page 15: ...ons as provided in the instructions for use of the disinfectant Never use a mix of disinfectants or a mix of disinfectants and soap Preferably apply the product to the imitation leather using a cloth...

Page 16: ...ny mechanical damage to the couch do not use the couch and contact your supplier Expected Service Life Provided the annual maintenance is carried out by a qualified Enraf Nonius service technician or...

Page 17: ...Length Depending on the configuration approximately 205 215 cm Height Adjustable from approximately 48 to 96 cm Weight Depending on the configuration approximately 100 kg Upholstery Biocompatible In a...

Page 18: ...cluding without limitation damages for loss of goodwill work or productivity computer malfunctions and any other form of business damage or loss even if advised of the possibility of their occurrence...

Page 19: ...Page 19 of 20 EN109 1570750 41 IFU...

Page 20: ...EN109 1570750 41 IFU Copyright Enraf Nonius B V Vareseweg 127 3047 AT Rotterdam The Netherlands Tel 31 0 10 20 30 600 info enraf nonius nl...

Reviews: