50
1600751-4D
•
Select S/D Curve
Triangular with the
central controller [5]
and enter with the
confirmation button [6]
•
Select the phase
duration with the
central controller [5],
you can start on
highest (1000 ms) or
lowest value (100 µs)
•
Push the confirmation
button [6] you activate
the intensity setting.
•
Use the central
controller [5] to adjust
the intensity.
•
Every push on the
special probe or every
beep will generate one
pulse.
•
Confirm the value by
pushing the confirmation
button [6]
•
Select the second the next phase duration with the central controller in the list. To receive a
correct calculation you need to select minimum 10 different phase durations, always include
500 ms.
•
Select “
” (left corner) to return to the previous menu.
Electrotherapy
2
Diagnostics
Type Electrode
Beep
Rheobase / Chronaxie
S/D Curve Rectangular
S/D Curve Triangular
Report
mA
Diagnostics
2
S/D Curve Triangular
1000 ms
700 ms
500 ms
300 ms
200 ms
100 ms
70 ms
5.0
mA