2.4: Battery Management: BQ20Z95R
The Mobius 5 has a battery management integrated circuit that monitors voltage of the cells,
temperature, and current going in and out of the pack. It uses the three measurements to
estimate state of charge and verify the battery is operating within its specifications. The
BQ20Z95R also controls bypass for the cells to keep the pack in balance. The nRF52832 can talk
to the BQ20Z95R over I2C to get details on status of the battery.
3.
Antenna Description:
The Mobius 5 contains a Molex 1461530300 flat patch antenna. The antenna is impedance-
matched to the RF output of the nRF52832 at 2.4 GHz with a 2450FM07A0029T matching
network. Maximum peak antenna gain of 2.5 dBi is calculated from measured e.i.r.p.
FCC (United States) Compliance Statement
The Mobius 5 does not have a user replaceable antenna. The only valid certification is with the
Molex 1461530300 patch antenna. Accessing the antenna connection and operating with a
different antenna is strictly prohibited.
Warning:
Changes or modifications not expressly approved by Enovate Medical could void FCC
compliance and negate the authority to operate the product.
IC (Canada) Compliance Statement
The Mobius 5 does not have a user replaceable antenna. This radio transmitter has been
approved by Industry Canada to operate only with the Molex 1461530300 patch antenna.
Accessing the antenna connection and operating with a different antenna is strictly prohibited.
(RSS-GEN clause 8.3)
Label (example)
SERIAL #
SERIAL # BREAKDOWN
MODEL
DATE YYMM
SERIAL #
311761720010001
3117617
2001
0001