background image

AM-5150

 

 

ENMET Corporation

 

12 

 
 

5.2 Calibration of the AM-5150 

Sensor Purge:

 Prior to calibration some 

MOS

(Metal Oxide Sensor) sensor should be purged.  A sensor purge is performed to 

clean the 

MOS

 sensor of molecules that may have settled on the sensor. 

Press and hold the 

S

ELECT

 switch, 3 to 5 seconds, if the purge feature is operational 0ppP appears in the display.  If the 

AM-

5150-MOS

 is equipped with the purge feature, the display will alternate between 0ppm and 0ppP and the green Power/Fault 

LED will flash.   

Example of purge display 

In approximately 2 minutes the purge function will be completed and the display will, alternate between 0ppm and 0ppR 
indicating the instrument is in recovery mode.  Recovery mode lasts approximately 5 minutes. 

 

Example of purge recovery display 

When the purge cycle is complete, the display will become stable at 0ppm and the green Power/Fault LED will stop flashing. 
Calibration is the process of setting the instrument up to read accurately when exposed to the target gas.  The Zero function sets 
the clean air reference point and the Span function sets the sensitivity of the instrument. 

Initial Calibration: 

Wait at least 3 – 4 hours after initially supplying power to the 

AM-5150

 instrument before initial 

calibration, overnight is preferred.  The 

AM-5150

 has been pre-calibrated at the factory, and initial field calibration should 

result in only fine-tuning to circuit, as well as a way to check that installation is successful.  It is not necessary to open the 
enclosure to make adjustment.  The calibration functions are operated with pushbuttons from outside the enclosure through the 

MENU

 and 

SELECT

 switches. 

Calibration Zero and Span functions are two separate procedures.  They operate independently of each other.  It is 
recommended that the Zero procedure be done prior to the Span procedure. 

ENMET

 Corporation recommends at least quarterly calibration of the 

AM-5150

 instrument

Calibration equipment is available from 

ENMET

 Corporation to calibrate the 

AM-5150

 instrument. 

  Calibration adapter, a length of tubing with a regulator for the gas cylinder on one end, and a calibration cup to connect to 

the sensor of the 

AM-5150

 on the other. 

  Gas cylinder, Zero gas 20.9% oxygen or Span gas, factory determined, varies by intent. 

Generally, a cylinder of 20.9% Oxygen is used to provide a Zero point or fresh air reference for the calibration.   

N

OTE

:

 Software revision may cause variations of display output. 

 

Figure 8: Calibration Adapter 

Menu

Select

Sensor 
Enclosure
 

Sensor 

Gas Cylinder 

Regulator 

Humidifier 

Calibration Cup 

To Cal Cup 

Top View of Humidifier 

To Regulator 

0ppm 

0ppR 

0ppm 

0ppP 

Summary of Contents for AM-5150

Page 1: ...ENMET Corporation PO Box 979 Ann Arbor MI 48106 0979 www enmet com Manual part number 80003 550 MCN 453 07 19 11 AM 5150 Ambient Air Monitor Manual...

Page 2: ...15 5 3 SENSOR REPLACEMENT 16 5 3 1 Sensor Replacement Calibration 17 6 0 ACCESSORY AND REPLACEMENT PARTS 18 7 0 TECHNICAL DATA AND SPECIFICATIONS 18 8 0 WARRANTY 19 List of Tables TABLE 1 RELAY FAILSA...

Page 3: ...T it legally became your company s property If your shipment arrives damaged Keep the items packing material and carton As Is Within 5 days of receipt notify the carrier s local office and request imm...

Page 4: ...ormation is displayed Audio Alarm Horn Audio alarm 105 dB at 30cm 12in The audio alarm is activated when the unit is in alarm Visual Indicators and Alarms LED indicators Power Fault Indicator LED Gree...

Page 5: ...ircuit Board For the contacts for each of three alarm relays and for the contacts of a fault relay See Section 3 3 Terminal J12 For VDC back up power in and the 4 20 mA output See Section 3 1 2 Sensor...

Page 6: ...f the installation to appropriate codes is not ENMET s responsibility The AM 5150 should be powered through circuit breakers provided for this purpose 3 1 2 Power Supply The input power can vary from...

Page 7: ...low Heavier than Air Gas Sensor Location Bottled LP liquefied petroleum Interior wall 18 24 from floor DO NOT locate directly above or beside gas appliances ovens heaters Avoid locating anywhere near...

Page 8: ...Heater Ground Brown 4 Signal Ground Yellow Wiring for 3 wire Terminal Sensors Position Function Wire Color 1 Heater Orange 2 Signal Blue 3 Heater Ground Brown Wire Nut Signal Ground Yellow NOTE The ye...

Page 9: ...ly Open J14 K1 Relay 1 NC Normally Closed J14 K1 Relay 1 COM Common Alarm 2 J15 K2 Relay 2 NO Normally Open J15 K2 Relay 2 NC Normally Closed J15 K2 Relay 2 COM Common Alarm 3 J16 K3 Relay 3 NO Normal...

Page 10: ...r is a flashing C The instrument has failed Calibration The last good calibration values are retained but the sensor may not be responsive to gas A new Calibration should be performed As Soon As Possi...

Page 11: ...l Scale Press MENU button Display returns to gas measurent Operational Display Flow Chart 4 2 1 Alarm Conditions AM 5150 There are three alarm set points available These alarm points are normally set...

Page 12: ...ay output TABLE 2 AM 5150 MAINTENANCE MENUS SEQUENCE Example of Display Function Normal Display Mode Measurement of CO Press and hold the MENU button for 2 4 seconds to enter the Maintenance Menu The...

Page 13: ...0 To change Alarm set points Press Menu button until Alarm to be changed is displayed Press Select button to display the set point The MENU button changes digit indicated by underscore cursor The SELE...

Page 14: ...5150 instrument before initial calibration overnight is preferred The AM 5150 has been pre calibrated at the factory and initial field calibration should result in only fine tuning to circuit as well...

Page 15: ...ll be made Cal OK appears on the display and in 1 2 seconds display will change to Span If you wish to Span the sensor press the SELECT button you are now ready to apply gas Proceed to gas span step 2...

Page 16: ...To exit cal press MENU button until Exit appears and press SELECT button If the sensor is outside of acceptable parameters Bad Span is displayed If the sensor did not respond an incompatible span gas...

Page 17: ...span range Enter the maintenance menu as shown in Figure 7 AM 5150 Maintenance Menu flow chart 1 Press the MENU button until to display Span is displayed 2 Press the SELECT button to initiate the mA S...

Page 18: ...s on side of sensor see Figure 9 Example 813 4 the first 3 digits are sensor ID the 4 digit indicates a 4 wire sensor 2 Disconnect the instrument for power 3 Disconnect the orange brown blue and yello...

Page 19: ...lt in only fine tuning to circuit as well as a way to check that installation is successful It is not necessary to open the enclosure to make adjustment The calibration functions are operated with pus...

Page 20: ...n calibration gases must be in a background of air do not use gases with an inert gas background such as nitrogen or argon 7 0 Technical Data and Specifications Electrical Power 15 Amp fused branch ci...

Page 21: ...product The purchaser indemnifies and saves harmless the company with respect to any loss or damages that may arise through the use by the purchaser or others of this equipment This warranty is expre...

Page 22: ...processing delays ENMET charges a one hour minimum billing for all approved repairs with additional time billed to the closest tenth of an hour All instruments sent to ENMET are subject to a minimum...

Page 23: ...nt Terms K COD Check one K VISA MasterCard______________________ ________ ________ Card number Expiration Card Code K American Express______________________ ________ ________ Card number Expiration Ca...

Reviews: