24
EQ-99X Operation Manual Rev. 2
The procedure below will explain the proper steps to appropriately assemble the water cooling
components of the EQ-99-CAL unit.
1.
The Swagelok fittings will need to be attached to the two copper tubing, as shown in
Figure 9.
a.
The ¼” end of the fitting will mate to the copper tubing.
b.
A 9/16” and 5/8” wrench will be required in order to complete the
attachment.
2.
The 3/8” end of the Swagelok fitting will connect to the two 3/8” OD Tubing, as
shown in Figure 9.
a.
A 5/8” and an 11/16” wrench will be required in order to complete the
connection.
3.
The open ends of the 3/8” tubing will connect to the “inlet” and “outlet” ports on the
300W Thermocube Chiller.
a.
The ports on the chiller will be “push-to-connect” fittings, so the tubing
should be pressed-in for a good seal.
b.
There are no designated ports for the two tubing. Either tubing can fit in any
port.
Figure 9:
Assembled EQ-99-CAL Lamp Head